OHIO WILLOW WOOD LimbLogic VS Series Prosthetist'S Manual Download Page 9

9

Charging the Vacuum Pump Battery

 

1.  Remove the cap from the charging port on the Vacuum Pump.

2.  Insert the charging plug into the charging port as follows:

a)  Gently insert the plug into the port, holding the black plastic  

portion of the plug, with just enough force to feel the plug touch 
the port.

b)  Align the green alignment marks on the charging port and the 

charging plug.

The charging plug can only be inserted into the Vacuum 
Pump charging port in one orientation. Forcing the plug will 
damage the plug, will possibly damage the vacuum system, 
and will void the warranty.

3.  Plug the charger into a power outlet. The 

blue

 light on the two-pronged 

adapter will light, followed by the 

green

 

LED.  The fob will display the 

battery charging

 indicator.

 

If the 

green

 

LED

 

does not light, press and hold the 

reset button

 until 

the light comes on.

4.  When charging is complete, the 

green

 

LED

 

will flash. Disconnect 

the power cord from the Vacuum Pump and from the power outlet.

charging port

Charger 

Vacuum Pump 

To 

AC 

outlet

Green LED

SOLID = charging

FLASHING = charged

charging plug

reset button

Align the 

green 

mark on the plug 

with the 

green 

mark on the port.

Overview

charger node

Summary of Contents for LimbLogic VS Series

Page 1: ...S Patient Information is included with each system and should be provided to the patient upon delivery of a prosthesis containing the LimbLogic VS The Vacuum Pump and fob included in this package are...

Page 2: ...as the only means of attachment of prosthesis to patient All LLVS components have passed safety testing for use as medical devices Radio enabled devices transmit with a power of less that one milliwat...

Page 3: ...on the Vacuum Pump 11 Turning on the Fob 12 Information Screen 13 Active Mode 13 For Bilateral Applications 14 Standby Mode 14 Configuration Screen 15 Detecting a Leak 16 If the Sealing Sleeve is Punc...

Page 4: ...oint press up and down buttons to adjust and center button to confirm Vacuum Pump battery charge level 4 To enter Standby Mode Press center button until appears Standby Mode Press center button until...

Page 5: ...he Vacuum Pump Battery on page 9 3 Properly secure the fully charged Vacuum Pump to the socket Vacuum Plate LLV 01043 or Vacuum Pyramid LLV 01044 with M6 flathead screws Tighten the screws to 9 ft lbs...

Page 6: ...iew Figure 1 Folding over the top edge of the liner Socket Alpha Flex Sleeve Fabric side of Alpha Liner Gel side of Alpha Liner BK Application The patient should don an Alpha Liner and Alpha Flex Slee...

Page 7: ...iner and Alpha AK Sleeve as shown below Figure 3 Top edge of the liner is folded over the brim of the socket Socket Gel side of Alpha Liner Gel side of Alpha Liner Fabric side of AK Sleeve Figure 4 Sl...

Page 8: ...value the Upper Set Point The difference between the Lower Set Point and the Upper Set Point is called the Vacuum Control Range The Upper Set Point cannot be set higher than the User Maximum The User...

Page 9: ...lug will damage the plug will possibly damage the vacuum system and will void the warranty 3 Plug the charger into a power outlet The blue light on the two pronged adapter will light followed by the g...

Page 10: ...should last for the life of the product Charging the Vacuum Pump Battery in a Vehicle To use a vehicle s power to charge the Vacuum Pump you must purchase a Power Inverter rated at 75 Watts or less t...

Page 11: ...the pump has just been unplugged from the charger the button does not need to be pressed When you are ready for the pump to begin regulating the vacuum press the on off button again or turn on the fo...

Page 12: ...left button Overview down button Press and hold the center button on the fob until the LimbLogic VS logo appears and the display has reached the desired brightness Release the button If the Vacuum Pu...

Page 13: ...displayed to the left of the battery symbol as shown below the system is in Active Mode In this mode the Vacuum Pump activates as needed to maintain the vacuum settings You or the patient may use the...

Page 14: ...RIGHT pump B Standby Mode Allows you to turn off the pump temporarily to prevent the pump from regulating the vacuum To enter Standby Mode press and hold the center button until the hourglass symbol a...

Page 15: ...adjust the User Maximum The User Maximum cannot be set higher than 20 in hg Press the up or down button to scroll to the next screen Use the right and left buttons to adjust the Vacuum Control Range T...

Page 16: ...tton on the fob until the hourglass symbol appears or Press the on off button on the Vacuum Pump one time If the Sealing Sleeve is Punctured The latex sleeves LLS 3N LLS 4N or LLS 5N in the Diagnostic...

Page 17: ...cuum Pump battery has less than 15 of its capacity remaining the Vacuum Pump enters a stutter mode Stutter mode is when the pump quickly turns on and off when pulling vacuum Charge as soon as possible...

Page 18: ...etone or similar chemicals to clean the system Please demonstrate this procedure for the patient before sending the patient home with the system Exposure to Water The Vacuum Pump is waterproof in fres...

Page 19: ...f a cosmetic cover The kit also includes a hose barb for routing the exhaust tube into the foot shell Vacuum Plate Pyramid The Vacuum Plate LLV 01043 and Vacuum Pyramid LLV 01044 can be used for attac...

Page 20: ...mp will not vibrate if it is already holding a vacuum when it is started up Do not press the on off button too quickly Press the button for one second then release for one second before pressing again...

Page 21: ...nition of the button activation Wait one minute for the fob screen to turn off Press the fob center button again The fob may be too close or too far from the pump Place the fob about 2 feet away from...

Page 22: ...tact Ohio Willow Wood With the charger plugged into an AC outlet and the pump press and hold the reset button on the charger node for 5 seconds then release If the green LED does not blink or turn sol...

Page 23: ...nother room The configurable prong may not be seated properly in the back of the charger s two pronged adapter Unplug the charger from the AC outlet Remove the charger s prong using the sliding latch...

Page 24: ...l the hole with epoxy then drill a new 9 32 hole in the center The leak was coming from the Vacuum Pump Contact Ohio Willow Wood If the leak does NOT go away If the leak goes away If the leak goes awa...

Page 25: ...Wood s recommendations and will void the two year warranty Extremely high and abusive activities are defined as activities such as skydiving karate and judo activities that could result in injury to...

Page 26: ...al property rights reflected or incorporated in its physical products regardless of the transfer of the physical products to another party or parties There are no field serviceable parts inside the Va...

Page 27: ...his product complies with IEC 60601 1 1998 A1 1991 A2 1995 EN 300 440 1 V1 3 1 2001EMC EN 301 489 1 V1 7 1 2006 EN 301 489 3 V1 4 1 2002 EMC and EN55022 1998 A1 2000 A2 2003 IEC 60601 1 Classification...

Page 28: ...ically tested for wound healing or other uses Ohio Willow Wood does not currently support the use of the LLVS for uses other than suspension Use only the bolts supplied by The Ohio Willow Wood Company...

Page 29: ...m Pumps use Ohio Willow Wood charger part number LLV 01010 Use with other products may result in harm to the device the user or both and void the warranty Refer to product manuals to verify compatibil...

Page 30: ...f the FCC Rules and RSS 210 Annex 2 9 This module has been designed for use by the Ohio Willow Wood Co This module shall only be designed into products by The Ohio Willow Wood Co Warning Changes or mo...

Page 31: ...12V DC Input current 37 mA Operating temperature range 20 C to 50 C Frequency range 2400 2483 MHz Antenna Permanently soldered to Module Transmit power 2mW Communication type Frequency Hopping Spread...

Page 32: ...etic environments specified below The customer or the user of the LLVS should assure it is used in such an environment Emissions Test Compliance Guidance RF Emissions Group 2 The LLVS must emit electr...

Page 33: ...kV Differential Mains power quality IEC 61000 4 5 Mode Mode should be that of a typical 2 kV Common 2 kV Common commercial or hospital Mode Mode environment Voltage Dips 5 UT 5 UT Mains power quality...

Page 34: ...Hz to 2 5 GHz IEC 61000 4 3 Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m...

Page 35: ...mated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80...

Page 36: ...s Pending OHIO WILLOW WOOD free the body free the spirit 15441 Scioto Darby Road Mt Sterling OH 43143 phone 740 869 3377 800 848 4930 fax 740 869 4374 www owwco com Ohio Willow Wood Company B V Keizer...

Reviews: