1. INSTRUCTIONS FOR USE 
 
PRIOR TO INITIAL USE 

 
Upon receiving your new Moblvac

®

 III-cs, perform the following initial tests to ensure that your unit is 

in good working order and that no damage has occurred during shipment. 
 

SET UP INSTRUCTIONS 

 
Your Moblvac

®

 III-cs has been shipped with some minor assembly required.  A 3/32” Allen wrench is 

the only tool required. 
 
1.  

Visually inspect all components for physical damage that may have occurred during shipping. 

 
2. 

After removing all components from their cartons, insert the backpole assembly into the back 
of the Moblvac

®

 III-cs base. 

 
3.  

Align the two “collection mounting brackets” located at the top of the backpole assembly to-
wards the front of the unit. 

 
4.  

Tighten the 10-32 set  screw located at the back of the base assembly with a 3/32” Allen 
wrench. 

 
5. 

Connect the tubing from the bottom of the backpole to the safety overflow trap assembly. 

 
6. 

Plug the power cord into an electrical outlet.  Depress the “POWER” switch and listen to verify 
that the pump starts. 

 
7. 

Plug the vacuum port located at the top of the backpole and adjust the vacuum level by turning 
the vacuum regulator knob.  Verify that the vacuum gauge reflects a charge in vacuum level 
while turning the regulator knob. 

 
8. 

Depress the “POWER” switch to turn the unit OFF (O). 

 
9.  Place a collection canister bracket (ring) in the slide mounting bracket of the backpole assembly or 

pedestal stand and insert a collection canister. 

 
 

IMPORTANT:  This device is for use only by persons properly trained in medical suction  

techniques and in the operations of suction equipment.  Improper use could cause injury.  

Thoroughly read this operations manual to familiarize yourself with the Moblvac

®

 III-cs  

before using the device. 

 

DANGER: POSSIBLE EXPLOSION HAZARD IF USED IN THE PRESENCE OF  

FLAMMABLE  ANESTHETICS.

 

Summary of Contents for Moblvac III-cs

Page 1: ...CONSTANT SUCTION UNIT OPERATOR MAINTENANCE M A N U A L Ohio Medical Corporation 1111 Lakeside Drive Gurnee Illinois 60031 4099 1 800 662 5822 847 855 0800 Fax 847 855 6218 www ohiomedical com...

Page 2: ...es Because of the Moblvac s constant suction capacities are driven by a fan cooled rotary carbon vane pump It comes equipped with a disposable exhaust filter an in line suction filter with 14 of tubin...

Page 3: ...w located at the back of the base assembly with a 3 32 Allen wrench 5 Connect the tubing from the bottom of the backpole to the safety overflow trap assembly 6 Plug the power cord into an electrical o...

Page 4: ...Moblvac III cs to the vacuum port of the collection canister and from the patient port of the collection canister to the aspirating tip that is being used 2 Open the vacuum port located at the top of...

Page 5: ...port located at the top of the backpole Adjust the Ohio Medical needle valve or Chest Drainage Unit per the manufacturers instructions as necessary and proceed with the suction procedure REPROCESSING...

Page 6: ...ng 2 Clean the overflow trap assembly by unscrewing the cap and disassembling the various compo nents Wash all parts with mild antiseptic dry reassemble and verify that the float is in the down positi...

Page 7: ...nections from the regulator body 3 Use a small flathead screwdriver to loosen the set screw on the vacuum regulator knob Un screw and remove the vacuum regulator knob from the control panel 4 Remove t...

Page 8: ...e regulator in the Moblvac III cs 8 Attach new vacuum tubing PUMP REPLACEMENT 1 Disconnect the three 3 pump wires from the PCB and two 2 wires from the capacitor 2 Disconnect the tubing from the fitti...

Page 9: ...he three 3 pump head screws 7 Reinstall the pump assembly 3 PERFORMANCE TEST AND ADJUSTMENT It is recommended that you verify performance or the Moblvac III cs after An overflow Preventative maintenan...

Page 10: ...the unit into an outlet 2 Faulty electrical connections 2 Make sure that all wires are secured tightly on the lugs and the lugs themselves are secure on the terminals 3 Pump has seized 3 Clean pump a...

Page 11: ...Hg ELECTRICAL REQUIREMENTS AC 120V 60Hz 2A 220V 50Hz 2A Fuses Slow blow COLLECTION DEVICE Canister Disposable plastic with mechanical shutoff Capacity 1200cc standard Tubing 14 with bacterial filter P...

Page 12: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 12 FIGURE 1...

Page 13: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 13 Figure 2...

Page 14: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 14 FIGURE 3...

Page 15: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 15...

Page 16: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 16 FIGURE 4...

Page 17: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 17 FIOGURE 5...

Page 18: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 18 FIGURE 6...

Page 19: ...Ohio Medical Corporation 1111 Lakeside Drive Gurnee Il 60031 4099 800 662 5822 Fax 847 855 6218 www ohiomedical com 550107 Rev 5 08 2005 19...

Page 20: ...ion 800 662 5822 Fax 847 855 6218 www aerosinstruments com 550107 Rev 5 08 2005 1111 Lakeside Drive Gurnee IL 60031 4099 Phone 800 662 5822 847 855 0800 Fax 847 855 6218 Email info ohiomedical com htt...

Reviews: