OHAUS VXHDAL Instruction Manual Download Page 9

8

A

CCESSORY

 I

NSTALLATION

 I

NSTRUCTIONS

 (A

LL

 U

NITS

)

Installing the cup head, microplate trays, universal holder or other head 
attachments:

1.  Move the 3-way rocker switch to the center, off/standby position. 

2.  Place accessory on head mount of the unit, aligning the square on the bottom of 

the accessory with the square on the top of the unit. Lower accessory until it is 
sitting on the head mount.

3.   Turn the accessory counter-clockwise until it snaps into place (approximately 

1/8 of a turn). Add samples to attachments after the accessory is secured onto 
the Vortex Mixer. Be sure samples are secure before turning unit on.

4.  To remove an installed accessory, move the 3-way rocker switch to the center, 

off/standby position. Turn the accessory clockwise (approximately 1/8 of a turn) 
and then lift the accessory straight up. Remove samples before removing the 
accessories. 

•   To load a microplate into the microplate tray, position the microplate against 

the foam bumpers first, then push down the other side to ensure a secure fit. 
To release a microplate, push it against the microplate tray’s foam bumper and 
lift to remove.

•   Maximum speed may vary when vortexing microplates. Speed may be 

dependant on the height of the skirt on the microplate or material of the 
microplate.

•   Rubber vibration dampeners on the bottom of the microplate tray and universal 

holder may need to be reinserted into the proper position when vortexing at 
high speeds. Squeeze firmly as you maneuver it back into position.

A

CCESSORY

 I

NSTALLATION

 I

NSTRUCTIONS

 (H

EAVY

 D

UTY

 V

ORTEX

 M

IXER

)

Installing the vessel holders:

1.  Attach the Universal Holder to the vortex mixer. 

2.  Place the Grip Mat in the bottom of the Universal Holder.

3.  Place the flask on top of the Grip Mat, in the center of the holder.

4.  Pull the center hole of the Vessel Holder over the sample container.

5.   With the arrow on the Vessel Holder close to the front of the unit, the straps 

should line up with the posts of the Universal Holder.

6.   Push the hole on the strap over the posts on the holder. Secure the strap so 

that it fits into the slot on the holder. Repeat this for the remaining two straps.

7.  Be sure the sample is secure before turning unit on. 

Installing the Universal Holder Cover:

1.  Place the Foam Insert into the Universal Holder.

2.  Lay the Universal Holder Cover on top of the foam, and align the edges. 

3.   Push the hole on the cover down over the post on the holder. Squeeze the 

strap so that it fits into the slot on the holder, leaving the cover sitting flat on 
the foam. Repeat this process for the remaining two posts.

4.  To remove the accessory cover, secure the straps and pull straight up.

EN

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for VXHDAL

Page 1: ...ex Mixer Heavy Duty VXHDDG Vortex Mixer Heavy Duty VXHDAL Vortex Mixer Microplate VXMPDG Vortex Mixer Microplate VXMPAL Revision 10 2 14 2020 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Page 2: ...ct service specialist will be available to provide assistance Outside the USA please visit our web site to locate the Ohaus office nearest you Serial Number _______________________________________________________ Date of Purchase ____________________________________________________ Supplier ___________________________________________________________ TABLE OF CONTENTS Package Contents 1 Service Inf...

Page 3: ...lastics nor one which is flammable Always ensure the power is disconnected from the unit prior to any cleaning If the unit ever requires service contact your Ohaus representative ENVIRONMENTAL CONDITIONS Operating Conditions Indoor use only Temperature 4 to 40 C 39 to 104 F Humidity 20 to 85 relative humidity non condensing Altitude 0 to 6 562 ft 2000 M above sea level Non Operating Storage Temper...

Page 4: ... 2 051 EN 61326 1 The full text of the EU declaration of conformity is available at the following internet address This product complies with directive 2012 19 EU Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment For disposal instructions in Europe refer to EN 61326 1 CAN CSA C22 2 61010 1 CAN CSA C22 2 610...

Page 5: ...lt quick acting Speed range On Mode 300 2500rpm Touch Mode 300 3500rpm Orbit 0 138 3 5mm Duty rating Continuous duty Controls Analog See page 4 Digital See page 5 Capacity 2 Microplates Ship weight 12lbs 5 4kg Overall dimensions L x W x H 9 5 x 6 6 x 6 3 24 1 x 16 8 x 16 5cm Electrical 50 60 Hz 120 volts 0 25 amps 30 watts 230 volts 0 13 amps 30 watts Fuses 5mm x 20mm 5 amp 250 volt quick acting S...

Page 6: ... how much time is remaining timed mode The display range is from 0 to 9 999 minutes in one 1 second increments The display will indicate minutes and seconds until the timer reaches 99 minutes and 59 seconds 99 59 then the display will automatically display minutes up to 9 999 C Up down arrows for set point control D Speed display Displays the speed of the Vortex Mixer E Up down arrows for set poin...

Page 7: ...he power cord into a properly grounded outlet 2 For continuous operation push the rocker switch to the right on position For intermittent touch operation push the rocker switch to the left touch position 3 In either case turn the speed knob to the desired setting In the on position you will see the head in motion immediately In the touch position the vessel you are using must be pushed down on the...

Page 8: ...down arrows below the time display until you reach the desired time While running unit in timed mode the time display will show time remaining counting down to zero 0 00 Four 4 audible beeps will indicate the count down function is complete The time display will default back to the set time To run in continuous mode accumulated time reset the timer to zero 0 00 before running Time display will sho...

Page 9: ...plate Rubber vibration dampeners on the bottom of the microplate tray and universal holder may need to be reinserted into the proper position when vortexing at high speeds Squeeze firmly as you maneuver it back into position ACCESSORY INSTALLATION INSTRUCTIONS HEAVY DUTY VORTEX MIXER Installing the vessel holders 1 Attach the Universal Holder to the vortex mixer 2 Place the Grip Mat in the bottom ...

Page 10: ...or or the motor is not working properly Analog and Digital Models 1 Turn the rocker switch to the center off standby position 2 Check to verify the load has not been exceeded 3 Check and clear any possible obstructions of the head mount 4 If the unit still does not work please contact your Ohaus representative EN www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Page 11: ...e o non funziona cor rettamente Modelli analogici o digitali 1 Girare la manopola oscillante al centro in posizione off standby 2 Controllare che non sia stato superato il carico massimo 3 Controllare ed eliminare qualsiasi ostacolo alla testa del dis positivo 4 Se l unità continua a non funzionare contattare il rappresent ante Ohaus IT www GlobalTestSupply com Find Quality Products Online at sale...