APPENDIX C SOFTWARE SERVICE TOOL INSTRUCTIONS
Valor™ 3000 Series Service Manual
C-1
Ohaus Corporation www.ohaus.com
APPENDIX C.
SOFTWARE SERVICE TOOL INSTRUCTIONS
The Software Service Tool (Part Number 80252509) is required when either a main PC Board
or a Load Cell is replaced in a Valor 3000 scale. It is used to re-configure the scale to its original
parameters in the case of a PCB replacement, or new parameters in the case of a Load Cell
replacement.
C.1 Hardware and Software Setup
1. First, check that the scale’s
A-Off
feature, under the
SETUP
menu, is set to
OFF
.
If this setting is left
ON,
the scale will shut off during configuration.
2. To connect the scale to a personal computer, use the hardware interface:
(See Figures C-1 and C-2.)
– RS232 Interface and Cable (PN 71147376 or 72206287)
– A Communications Service Interface (PN 72203450) that attaches to the
RS232 Interface
3. Plug the cable into the PC’s RS232 port.
Figure C-1. RS232 Interface
connected to Service Interface,
connected to the scale’s PCB,
circuit board side up.
Figure C-2. Service Interface fits into the RS232
Interface with the gap on the left side when the label
side is up on the RS232 Interface.
4. Install the software on a Personal Computer running Microsoft Windows NT 4.0 or later,
or Microsoft Windows 98 or later.
5. Run the program Service Tool. (See next page.)
Summary of Contents for Valor 3000
Page 2: ......
Page 4: ......
Page 8: ...TABLE OF CONTENTS Valor 3000 Series Service Manual iv Ohaus Corporation www ohaus com...
Page 16: ...CHAPTER 1 GETTING STARTED Ohaus Corporation www ohaus com 1 8 Valor 3000 Series Service Manual...
Page 51: ......
Page 52: ...80252591 P N 80152591 SERVICE MANUAL Valor 3000 SERIES SCALES...