Traveler
™
Scales
EN-6
4.2.1 Setup Menu
[
S.E.t.U.P
S.E.t.U.P
S.E.t.U.P
S.E.t.U.P
.
]
The Setup menu contains the following sub-menus:
Auto-Off
[
.AAAA----OFF.
OFF.
OFF.
OFF.
]:
Auto Off settings:
On Off
Off
Off
Off
Setting [
On
] will automatically turn the scale off after no activity
is detected for 4 minutes.
Span Calibration
[
SPAN
SPAN
SPAN
SPAN
]:
Uses two weight values: Zero and a value at
or near the
full capacity of the scale
.
Span calibration should be
performed if the scale is repositioned or when the room
temperature changes significantly.
Linearity Calibration
[
LLLLin
in
in
in
]:
Uses three weight values: Zero, a mid-range value and full
capacity. Generally this calibration is not required unless
testing shows that the linearity error exceeds the linearity
tolerance in the specification table. Accuracy of weights is
important to maintain weighing integrity.
4.2.2 Unit Menu
[
U.n.i.t
U.n.i.t
U.n.i.t
U.n.i.t
]
Each Weighing Unit may be individually set to
On
or
OFF
.
See Table 4-1 for the symbols associated with each Unit or Mode.
4.2.3 Other Menus
Additional Menus are available if a serial interface option is installed in the scale. These
menus are explained within the user manual of each option kit.