CHAPTER 1
GETTING STARTED
1.9.3 Linearity
Calibration
Linearity calibration uses three calibration points. The full calibration point is established with a
weight on the scale. The mid-calibration point is established with a weight equal to half of the full
calibration weight on the scale. The zero calibration point is established with no weight on the
scale. The mid-calibration points cannot be altered by the user during the calibration procedure.
Press and hold
TARE/Menu
until
C.A.L.
appears, then press
Yes
.
(
LFt.NO
appears briefly before
C.A.L.
, if the Indicators is not in LFT mode.)
ZERO
appears. . Press
No
.
SPAN
appears. Press
No
.
Linearity
appears.
Press
Yes
to initiate Linearity Calibration.
The display flashes the full calibration point and calibration unit.
Place the specified weight on the scale and press
Yes
.
To choose a different full point or calibration unit (kg or lb), press
No
to increment the
selections, or
Back
to decrement the selections. When the desired value is displayed,
place the specified weight on the scale and press
Yes
.
The display shows
--C--
while the full calibration point is established.
The display flashes the mid-calibration point. Place the specified weight on
the scale and press
Yes
.
The display shows
--C--
while the mid-calibration point is established.
The display flashes
0
. With no weight on the scale, press
Yes
to establish the
zero point.
The display shows
--C--
while the zero point is established.
When linearity calibration is completed, the display shows
dONE
.
The scale exits to the active weighing mode and displays the actual weight
value.
Ohaus Corporation www.ohaus.com
1-12
5000 Series T51 Indicator Service Manual