OHAUS STARTER 300 Instruction Manual Download Page 12

 

10 

4.3 

Set manual temperature compensation 

After confirming the temperature unit, the MTC temperature appears on the display.  Then use 

up

 

or 

down

  buttons  to  increase  or  decrease  the  value  to  the  temperature  of  your  sample.  Press 

Read

 to confirm the setting. Press 

Exit

 to leave to setup. The default setting is 25 °C(77

°

F).  

Note:    

°

C = 5/9 (

°

F -32) 

4.4 

Selecting a predefined buffer group 

After  confirming  the  compensation  temperature  value,  it  goes  to  the  buffer  group  selection.  Use 

up

 or 

down

 buttons to select a buffer group(3.1.1). Press 

Read

 to confirm the setting. Press 

Exit

 

to leave to setup. The default buffer group is 

b1

.  

b1 

1.68 

4.01 

7.00 

10.01 

 

 (at 25°C) 

 

Maintenance 

5.1 

Error message 

Error 0 

Memory access error 

Reset to factory settings 

Error 1 

Self-diagnosis failed 

Repeat the self-diagnosis procedure and make 
sure that you finish pressing all five keys within 
two minutes.  

Error 2 

Measured values out of range 

Check if the electrode is properly connected and 
placed in the sample solution. 

Error 3 

Measured buffer temperature 
out of range (5 ... 40 °C) 

Keep the standard buffer temperature within the 
range for calibration 

Error 4 

Offset out of range 

Make sure the buffer is correct and  fresh; Clean 
or replace the pH electrode. 

Error 5 

Slope out of range 

Make sure the buffer is correct and  fresh; Clean 
or replace the pH electrode. 

Error 6 

Meter cannot recognize the 
buffer 

Make sure the buffer is correct and  fresh; check if 
the buffer has not been used more than once. 

Error 9 

The current data set has 
already been stored once 

A measurement can only be stored once. Perform 
a new measurement to store a new data set. 

5.2 

Meter maintenance 

Never unscrew the two halves of the housing! 
The STARTER 300 series instruments do not require any maintenance other than occasional wipe 
with a damp cloth and the replacement of depleted batteries. 
The  housing is made  of  acrylonitrile  butadiene styrene  (ABS).  This material  is  attacked  by some 
organic solvents, such as toluene, xylene and methyl ethyl ketone (MEK). Any spillage should be 
immediately wiped off. 

5.3 

Electrode maintenance 

Make sure the electrode is filled with electrolyte solution. Always store the electrode according to 
the manufacturer

’s instructions and do not allow it to dry out. 

 
If the electrode response becomes sluggish or the slope is not good enough, try: 
1. Soak the electrode in 0.1M HCl overnight 
2. For fat or oil build-up, degrease the membrane with cotton wool soaked in either acetone or a 
soap solution 

Summary of Contents for STARTER 300

Page 1: ...STARTER 300 Portable pH Meter Instruction Manual...

Page 2: ......

Page 3: ...libration 7 3 1 3 Performing 2 point calibration 8 3 2 Sample measurement 8 3 2 1 pH measurement 8 3 2 2 mV measurement 8 3 3 Temperature measurement 8 3 4 Using the memory 9 3 4 1 Storing a reading 9...

Page 4: ...permanent installations Integrated labeling design for quick product identification IP54 protection against dust and water Ohaus renowned user friendly software 1 1 Safety measures Measures for your...

Page 5: ...endpoint 3 Measurement icon measurement or calibration is running 4 Calibration icon calibration in progress 5 Setup icon instrument is in the setup mode 6 Battery status icon shows the condition of t...

Page 6: ...Start calibration Review the latest calibration data Meter on Back to measurement screen Meter off Store current reading to memory Increase value during setting Scroll up through the memory Recall sto...

Page 7: ...the batteries 1 Remove the battery cover screw and remove the battery cover 2 Insert the batteries in the battery compartment as shown 3 Replace the battery cover and tighten the battery cover screw...

Page 8: ...p forward to lock in position Slide the shaft of the sensor into the clip from above 2 4 Installing the IP54 cover Use the small plastic tool as shown 2 5 Integrated stand for table top use One unique...

Page 9: ...libration buffer and press Cal Cal 1 is blinking during calibration The calibration icon and the measurement icon appear on the display The measurement icon is blinking during calibration measurement...

Page 10: ...is recommended If you use the MTC mode you should enter the correct temperature value and keep all buffer and sample solutions at the set temperature see 4 2 To ensure the most accurate pH measuremen...

Page 11: ...confirm the deletion or press Exit to return to the measurement mode without deleting the data 4 Setup 4 1 Set endpoint mode The automatic endpoint is the default setting of the meter By pressing and...

Page 12: ...t out of range Make sure the buffer is correct and fresh Clean or replace the pH electrode Error 5 Slope out of range Make sure the buffer is correct and fresh Clean or replace the pH electrode Error...

Page 13: ...e to press the other keys until all the icons have disappeared When the self diagnosis has been completed successfully PAS appears If self diagnosis fails error message Err 1 appears Note You have to...

Page 14: ...9 03 10 28 50 1 98 4 06 6 97 8 99 10 10 Buffer group b3 JJG119 Temp pH1 68 pH4 00 pH6 86 pH9 18 pH12 46 5 1 67 4 00 6 95 9 39 13 21 10 1 67 4 00 6 92 9 33 13 01 15 1 67 4 00 6 90 9 28 12 82 20 1 68 4...

Page 15: ...Corporation 7 Campus Drive Suite 310 Parsippany NJ 07054 USA Tel 973 377 9000 Fax 973 944 7177 With offices worldwide www ohaus com PN 83033978 Ohaus Corporation 2011 all rights reserved 83033978 121...

Reviews: