EN-8
STARTER 2100 Bench pH Meter
3
SETUP
3.1
Set MTC temperature
Please note:
If a temperature electrode is used, Automatic Temperature Compensation (
ATC)
and the sample
temperature are displayed on the screen. You may then choose to skip MTC setup (below).
If the meter does not detect a temperature electrode or one is not used, the meter automatically
switches to Manual Temperature Compensation (
MTC
) mode and
MTC
appears on the screen.
MTC can be set as follows:
Press button-
and the MTC temperature value appears on the display.
Next use button-
or button-
to increase or decrease the value to the
temperature of your sample.
Press button-
to confirm the setting then return to the measurement screen; or
press button-
to reject the setting and return to the measurement screen.
The default MTC temperature value setting is 25 °C (77
°
F).
Note:
°C = 5/9 (°F -32)