CHAPTER 1 GETTING STARTED
Ohaus Corporation www.ohaus.com
1-6
Starter Series Bench Meter Service Manual
1.7 METER OPERATION – STARTER 2100
This section contains information on the basic operation of the Starter 2100 meter.
1.7.1 OVERVIEW OF THE CONTROLS
Displays
Figure 1-1. Starter 2100 display
1 Electrode condition
Slope: more than 95%
and offset: ± (0-15) mV
Electrode condition is
good
Slope: 90-95%
or offset: ± (15-35)
mV
Electrode condition is
acceptable
Slope: less than 90%
or offset: ± (35-60) mV
Electrode condition is not good
or needs cleaning
2 Measurement icon - ; measurement or calibration is running
3 Calibration icon -
Ca
l ; 1 point or 2 point calibration in progress
4
pH/mV
reading or slope (%) in calibration
5 Error message-
Err
6 Auto temperature compensation -
ATC
; Manual temperature compensation -
MTC
7
Temperature during measurement of
Offset
(mV) in calibration
Summary of Contents for ST3100M
Page 27: ......
Page 28: ...30332138B P N 30332138B SERVICE MANUAL Starter Bench Top Series...