CHAPTER 3 RESET TO FACTORY SETTINGS
Starter Series Portable Meter Service Manual 3-1
Ohaus Corporation www.ohaus.com
3.1 Recover factory settings
For ST300, ST300C and ST300D proceed as below:
When the meter is off, press and hold button-
Read
& button-
Cal
& button-
Exit
simultaneously for 3 seconds, the screen displays
and blinks,
means “Reset”. Then
we have 2 choice:
Press button-
Read
to reset factory settings (MTC, slope and offset, etc.).
Or press button-
Exit
to quit the setting and turn off the meter.
For ST400D proceed as below:
Switch ON the meter.
Press button-
and button-
simultaneously to enter the Reset Factory Setting
screen.
Meter screen will show as below where respective information is back to Factory Default.
Figure 3-1. ST400D Factory Default Screen.