OHAUS Scout SPX123 Instruction Manual Download Page 20

 

SCOUT

TM

 SERIES BALANCES - SPX                                                                                              EN-19 

 

Important notice for verified weighing instruments in the EU 

When the instrument is used in trade or a legally controlled application it must be set up, verified 
and sealed in accordance with local weights and measures regulations. It is the responsibility of 
the purchaser to ensure that all pertinent legal requirements are met. 
Weighing Instruments verified at the place of manufacture bear the following supplementary 
metrology marking on the descriptive plate. 

 

Weighing Instruments to be verified in two stages have no supplementary metrology marking on 
the descriptive plate. The second stage of conformity assessment must be carried out by the 
applicable weights and measures authorities. 
If national regulations limit the validity period of the verification, the user of the weighing 
instrument must strictly observe the re-verification period and inform the weights and measures 
authorities 
As verification requirements vary by jurisdiction, the purchaser should contact their local weights 
and measures office if they are not familiar with the requirements. 

 

Disposal 

 

In conformance with the European Directive 2002/96/EC on Waste Electrical and 
Electronic Equipment (WEEE) this device may not be disposed of in domestic 
waste. This also applies to countries outside the EU, per their specific 
requirements. 
 

Please dispose of this product in accordance with local regulations at the collecting point 
specified for electrical and electronic equipment. If you have any questions, please contact 
the responsible authority or the distributor from which you purchased this device. 
Should this device be passed on to other parties (for private or professional use), the content 
of this regulation must also be related. 
Disposal instructions in Europe are available online at www.ohaus.com/weee.  
Thank you for your contribution to environmental protection. 

 
FCC Note 

This equipment has been tested and found to comply with the limits for a Class B digital device, 
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable 
protection against harmful interference in a residential installation. This equipment generates, 
uses and can radiate radio frequency energy and, if not installed and used in accordance with 
the instructions, may cause harmful interference to radio communications. However, there is no 
guarantee that interference will not occur in a particular installation. If this equipment does 
cause harmful interference to radio or television reception, which can be determined by turning 
the equipment off and on, the user is encouraged to try to correct the interference by one or 
more of the following measures: 
-   Reorient or relocate the receiving antenna. 
-   Increase the separation between the equipment and receiver. 
-   Connect the equipment into an outlet on a circuit different from that to which the receiver is   
    connected. 
-   Consult the dealer or an experienced radio/TV technician for help. 
 

Industry Canada Note 

This Class B digital apparatus complies with Canadian ICES-003. 
 
 

(866) 433-6682 • (281) 359-8538 • [email protected] • www.novatech-usa.com

Summary of Contents for Scout SPX123

Page 1: ...Scout TM Series Balances SPX Instruction Manual sales novatech usa com www novatech usa com Tel 866 433 6682 Fax 866 433 6684 Tel 281 359 8538 Fax 281 359 0084...

Page 2: ...comply with these warnings could result in personal injury and or property damage Retain all instructions for future reference Verify that the local AC power supply is within the input voltage range p...

Page 3: ...ocation For best performance the Scout SPX balance should be used in a clean stable environment Do not use the balance in environments with excessive drafts with rapid temperature changes near magneti...

Page 4: ...mpartment Note After power on it is recommended to warm up the balance for at least 5 minutes before using it 2 5 Initial Calibration When the Balance is first installed and when it is moved to anothe...

Page 5: ...o Off Turns the balance Off Units Changes the weighing unit Mode Allows changing the application mode Menu Enter the User menu Menu Function Short Press Yes Accepts the current setting on the display...

Page 6: ...nd hold the On Zero Off button until OFF is displayed 3 3 Calibration Operation When the balance is operated for the first time a span calibration is recommended to ensure accurate weighing results Be...

Page 7: ...al pieces until the desired count is reached 8 To clear the stored APW press and hold Mode until Count is displayed Press Yes when Clr pwW is displayed Note Press Function to view the current APW 3 6...

Page 8: ...der icon will light If the sample is within the target weight range the accept symbol will light If the sample is over the target weight range over icon will light Note Press Function to view the low...

Page 9: ...clear the stored total press and hold Mode until total is displayed When Clr tot is displayed press Yes 3 9 Hold Mode There are two modes for the display hold Peak Hold allows the user to capture and...

Page 10: ...Peak Off Reset g kg ct N oz ozt dwt lb lb oz grn hkt sgt twt t tola tical C End End End End Notes Only available in certain regions When the Security Switch is in locked position the Calibration funct...

Page 11: ...hange of 3 divisions per second has been exceeded Stable Range Stable Set the amount the reading can vary while the stability symbol remains on 0 5d 0 5 balance division 1d 1 balance division 2d 2 bal...

Page 12: ...available for use with the Mode button Reset reset Reset the Mode menu to factory defaults NO not reset YES reset Weigh wWEIGH Set the status OFF disabled ON enabled Count COUNt Set the status OFF di...

Page 13: ...ts in the menu must be turned on to be active Note Available units vary by model and local regulations Custom Unit C Use the Custom Unit to display weight in an alternative unit of measure The custom...

Page 14: ...2 rESEt will be displayed Press No Unit g will be displayed 3 Press No to toggle through active units until Unit c is displayed 4 Press Yes On c will be displayed then press Yes 5 Factor is displayed...

Page 15: ...is level and secure and that the transportation lock has been released Power on the balance then use a string or wire to attach items to be weighed Figure 4 3 Weigh below application Connecting the I...

Page 16: ...ght 1d Balance shows error then exits parts counting or goes to Clr APU Lo ref Percent Reference weight or APW is too low for accurate results CAL E Fail to do calibration USB Err Cannot find menu or...

Page 17: ...ration Digital with external weight Tare Range Full Capacity by subtraction Weighing Units g kg ct N oz ozt dwt lb lb oz grn Tael HongKong Tael Singapore Tael Taiwan tical tola C Custom unit Applicati...

Page 18: ...200 g 300 g 2 kg 5 kg 8 kg Linearity Calibration Mass 200 400 g 300 600 g 1 kg 2 kg 3 kg 6 kg 4 kg 8 kg Stabilization Time s 1 Construction ABS plastic housing with 304 stainless steel SST pan Drafts...

Page 19: ...0 0005 4 8500 x 0 0005 13 6685 x 0 0005 18 080 x 0 005 Pound Ounce lb oz 1lb 5 870oz x 0 005oz 4lb 13 600oz x 0 005oz 13lb 10 700oz x 0 005oz 18lb 1 25oz x 0 05oz Grain grn 6480 x 2 9570 x 2 33950 x...

Page 20: ...or electrical and electronic equipment If you have any questions please contact the responsible authority or the distributor from which you purchased this device Should this device be passed on to oth...

Page 21: ...es to be defective at no charge provided that the product is returned freight prepaid to OHAUS This warranty does not apply if the product has been damaged by accident or misuse exposed to radioactive...

Page 22: ...ltweite Gesh ftsstellen Con uffici in tutto il mondo www ohaus com 30268991 P N 30268991 C 2016 OHAUS Corporation all rights reserved todos los derechos reservados tous droits r serv s Alle Rechte vor...

Reviews: