A-6
APPENDIX A SERVICE TOOL INSTRUCTIONS
A.6
USING CONFIG TOOL AFTER REPLACING THE MAIN PC BOARD (CONT.)
5.
The balance displays
Err 9.
6.
Start the ConfigTool software, see A.5
7.
An error message will be displayed
User Enable Units & User Variable Check Sum
Verify Error,
click on
OK.
8.
Set RS232 to the correct com port, Baudrate = 2400, ByteSize = 7, Parity = None.
9.
The balance will respond to the V command but not the others.
10. The bar graph will respond to pressure on the platform.
11. Enter the balance serial number.
12. Click on
OK
.
13. Click on
READ
.
14. The balance should read the correct model and serial number.
15. Perform the factory calibration. See B.2
16. The unit is finished. Remove the RS232 option or the jumper on the customers RS232
option.