OPERATION  

POWER On/Off:

 

To turn scale on, Press 

ON/OFF

. Scale conducts a display 

test. When the weight display appears, scale is ready to weigh. 

To turn scale off, press 

ON/OFF

OFF

 

appears briefly and display goes off. 

 

Zero Operation 

Zeroing corrects the influence of slight changes on the weighing pan. 

Manual:  

Unload scale and press 

ZERO

. The zero display appears. 

Automatic:  

As standard, the zero point of the scale is automatically corrected 

when scale is unloaded. (Automatic zero point correction is a menu setting.)

 

 

Weigh

 

Place item to be weighed on the pan. The display indicates gross weight  
in the selected measuring unit.   

Tare:

 Place an empty container on the pan and press 

TARE

.  

The Net weight and NET indicator are displayed. 
To clear Tare value, remove container and press 

TARE

. (Or press 

Clear

.) 

 

Auto-Tare: 

If 

A-tArE

 is activated in the menu 

(see next page)

, when the 

symbol flashes,  

place container on scale. Its weight is automatically tared. Zero and 

NET 

appear, and display 

shows weight of contents when placed in container. Tare weight is cleared automatically when 
scale is unloaded. 

Chain Tare: 

(activate 

CHAIn.tr

 

in menu) 

Tare several times: Place first container on scale, press 

TARE

. Zero and 

NET 

appear. 

Weigh sample, record result. Place second container on scale, press 

TARE

 again.  

Total weight on scale is saved as new tare weight; zero display appears.  
Add new sample to weigh, and record result. Repeat as needed. 

ADDITIONAL OPERATIONS

 

Changing Unit of Measure

  

Press 

Units 

to toggle between units available for active use.  

To add units, press 

MENU

 and go to 

Display

 to activate other available units: 

lb

 – pounds, 

oz

 – ounces, 

kg

 – kilograms, 

g

 – grams, or 

t

 – taels.  

(See back page for help using the Menu. Use Unit-Roll menu for access to all units.)

 

 

 

Dynamic Weighing 

(Note mouse graphic next to battery symbol.)

 

Use this setting to weigh moving samples, like live animals.   
Scale calculates mean value of 56 weighing operations in 4 seconds. 
Use the 

Applications/Average 

menu to choose either 

Manual

 or 

Auto

:  

 

Manual:

 Place sample on scale; when ready, 

press 

PRINT

. Dynamic weighing begins.  

Auto

: Place sample on scale;  

dynamic weighing starts automatically.  

During dynamic weighing, horizontal segments appear in the display.  
Weighing result (average of 56 readings) is displayed with the symbol 

*

Note: 

 To reset to normal weighing when finished with dynamic weighing: 

Press 

Menu 

then go to the 

Applications/Average 

menu, and choose 

OFF

, then 

Reset

 

Reviews: