EN-2 

2

nd

 Scale Interface 

SETUP 

After the main platform is connected to the 2

nd

 weighing platform, switch on the Dual Scale (D.Scale) 

menu item in the SETUP menu (see menu structure below).

  

Then set the CAP2 and Grad2 in the SETUP menu and do a calibration. Press and hold the 
key to switch between the main platform and the secondary platform.

 

CAL 

SETUP 

Zero2
Span1 
Span2
Lin1

 

Lin2

 

GEO 
End 

Reset 

D.Scale
Cap2
Grad2

Pwr. Un
Zero
A.Opt
A.Tare
Bp.Opt
Bp.Sig
Bp.Key
Accum
End

Note: 

Make sure to perform a calibration

 

of the scale before using the 2

nd

 weighing platform. 

Make sure the weighing unit is in kg or lb when using high capacity platform. (If display shows 
error8.6, long press [

Units

] key to switch unit to kg or lb.) 

ACCESSORIES 

For a complete listing of Ohaus printers and other accessories, contact Ohaus Corporation or visit 
www.Ohaus.com 

COMPLIANCE 

This accessory has been tested and complies with the approvals listed in the applicable Instruction 
Manual. 

Disposal 

In conformance with the European Directive 2002/96 EC on Waste Electrical and 
Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. 
This also applies to countries outside the EU, per their specific requirements. 
Please dispose of this product in accordance with local regulations at the collecting 
point specified for electrical and electronic equipment. If you have any questions, 
please contact the responsible authority or the distributor from which you purchased 
this device. 
Should this device be passed on to other parties (for private or professional use), the 
content of this regulation must also be related. 
For disposal instructions in Europe, refer to www.ohaus.com/weee.
Thank you for your contribution to environmental protection. 

Reviews: