EN-8 

PR Series Balance 

 

4.  APPLICATIONS 

The PR balance can be operated in 3 application modes by long pressing the 

Function / Mode

 button. 

 

4.1  Weighing 

Note: 

Before using any application, be sure the balance has been leveled and calibrated.   

Use this application to determine the weight of items in the selected unit of measure.    

 

Weighing 

 

1.  Press 

Tare

 or 

Zero

 if necessary to begin.   

2.  Press and hold the 

Function / Mode

 button to select 

wWEIGH

 

(this application is the default).   
 

 
 
 
 

 

 

3.  Place  objects  on  the  pan  to  display  the  weight.  Once  the 

reading is stable, the * will appear. 

4.  The resulting value is displayed in the active unit of measure. 

 

 

 

Item Settings 

 

 

To view or adjust the current settings.

 

 

 

Weighing Units:

 Change the displayed unit. See Section 5.4 for the detailed processes. 

 

Filter Level:

 Change Filtering level. See Section 5.3.1 for more information. 

 

GLP Data:

 See Section 5.7 for more information. 

 

Print Settings:

 Change printing settings. See Section 7 for more information. 

 

4.2  Parts Counting 

Note:

  Before  using  any  application,  be  sure  the  balance  has  been  leveled  and  calibrated.  The minimum  piece 

weight should be no less than 0.1d. 

Use this application to count samples of uniform weight.    
 

Parts Counting 

1.  Press 

Tare

 or 

Zero

 if necessary to begin. 

2.  Press and hold the 

Function / Mode

 button until 

Count 

appears. 
 

 

Summary of Contents for PR124

Page 1: ...OHAUS Pioneer PR Series User Guide...

Page 2: ......

Page 3: ...3 5 1 Menu Navigation EN 13 5 1 1 Changing Settings EN 13 5 2 Calibration EN 14 5 2 1 Calibration Sub menu InCal models EN 14 5 2 2 Internal Calibration not applicable to ExCal models EN 14 5 2 3 Cal...

Page 4: ...27 6 1 Settings EN 27 6 2 Verification EN 27 6 3 Sealing EN 27 7 PRINTING EN 28 7 1 Connecting configuring and Testing the Printer Computer Interface EN 28 7 2 Output format EN 29 7 3 Printout Exampl...

Page 5: ...eful information about the product Warning Symbols General Hazard Electrical Shock Hazard 1 4 Safety Precautions CAUTION Read all safety warnings before installing making connections or servicing this...

Page 6: ...e is level before it is used or after its location is changed The PR balance has a level bubble in a small round window beside the display To level the balance adjust the 2 leveling feet until the bub...

Page 7: ...tion When the PR balance is first installed or when it is moved to another location it must be calibrated to ensure accurate weighing results PR balances are classified into two categories InCal model...

Page 8: ...re operation Secondary Function Press and Hold Off x Zeroing current value Unit x Changes weighing units Mode x Changes application mode Menu Cal x Enters the main menu Calibration is the first sub me...

Page 9: ...its net weight is displayed After the container and the objects are removed the load will be displayed as a negative number Press Tare to clear Zero Press Zero to zero the balance 3 3 Overview of Part...

Page 10: ...ill appear 4 The resulting value is displayed in the active unit of measure Item Settings To view or adjust the current settings Weighing Units Change the displayed unit See Section 5 4 for the detail...

Page 11: ...ue to 15 please press No Then Put and 15 will flash simultaneously 4 Place 15 samples on the pan Press the Function Mode button so that the weight of the 15 samples is used to establish the average pi...

Page 12: ...ly as parts are added APW Optimization occurs only when the number of pieces added to the pan is between one and three times the number already on the pan Print Settings Changing printing setup See Se...

Page 13: ...is stored in memory The display will show 100 6 Remove the reference sample and place the test object on the pan The ratio of the test object to the reference sample weight is displayed as a percentag...

Page 14: ...eled and calibrated The PR balance is equipped with a weigh below hook for weighing below the balance as shown below Before turning the balance over remove the pan and draft shield elements if present...

Page 15: ...and Time Project Name Reset Time Format Ounce Troy Balance ID Time Setting Penny Weight Balance Name Brightness Newton User Name Auto Dim Grain Project Name LFT Application Name Result Gross Weight Ne...

Page 16: ...span calibration value on the pan and note the difference in divisions between the nominal mass value and the actual balance reading If the difference is within division calibration adjustment is not...

Page 17: ...e weights and 100 000 g place weight s of 100 g on the pan for calibration 5 Once the span calibration is completed successfully CAL Done will display Press any button to return to the previous screen...

Page 18: ...calibration mass value will be shown in the display After the display shows 100 0000 g please place weight s of 100 g on the pan for calibration 5 Remove the weight of 100 g from the pan After a whil...

Page 19: ...5 3 2 AZT Auto Zero Tracking Set the automatic zero tracking functionality Auto Zero Tracking Off disabled 0 5d display maintains zero up to a drift of 0 5 graduation per second 1d display maintains...

Page 20: ...e Format Set the current date format Date Format YY MM DD MM DD YY DD MM YY Date Format MM DD YYYY DD MM YYYY YYYY MM DD 5 3 6 Date Setup Set the current date in the desired date format To set the cur...

Page 21: ...m High Low Brightness Medium High Low 5 3 10 Auto Dim Set whether the balance automatically turns off the display backlight of the display Off disabled 10 minutes become dim if there is no motion for...

Page 22: ...ked at current setting x Graduations are forced to 1 Division and the menu item is hidden Communication Menu Communication Print Settings Print Output x Stable Weight Only is locked On x Numeric Value...

Page 23: ...ss No to advance to the next unit 3 Press Yes to set the unit displayed to the weighing unit 5 5 RS232 Interface Setup Enter this sub menu to customize RS232 standard settings Data may be output to ei...

Page 24: ...data bits odd parity stop bit 1 7 O 2 7 data bits odd parity stop bit 2 Parity 8 data bits stop bit 1 5 5 3 Handshake Set the flow control method NONE no handshaking XON XOFF XON XOFF handshaking HARD...

Page 25: ...2 On Stability printing occurs when the stability criteria are met When On Stability is selected set the conditions for printing Load Prints when the displayed load is stable Load and Zero Prints when...

Page 26: ...e Name On the balance name is printed Off the balance name is not printed 5 6 9 User Name On the user name is printed Off the user name is not printed 5 6 10 Project Name On the project name is printe...

Page 27: ...g 4 Lines move the paper up four lines after printing 1 Line 4 Lines 5 7 GLP Enter this menu to set the Good Laboratory Practices GLP 5 7 1 Header Enables the printing of GLP headings There are up to...

Page 28: ...available for each Header setting The default is blank 5 8 Factory Reset Use this sub menu to reset the all menu settings to their Factory default settings Reset All resets all menus to their factory...

Page 29: ...ode to On in the Balance Setup menu 5 As shown in Figure 6 1 press the push button inside the hole Note When Approved Mode is set to On external calibration can t be performed 6 2 Verification A weigh...

Page 30: ...ted Send line ends Echo typed characters Wrap lines Use RS232 Interface Commands Section 9 6 1 to control the balance from a PC SPDC Software The Serial Port Data Collection SPDC software is provided...

Page 31: ...haracters field contains CRLF Four CRLF or Form Feed ASCII 12 depending on the LINE FEED menu setting 8 When Numeric Only is set On only the Weight Field is printed left aligned 7 3 Printout Examples...

Page 32: ...erified By ____________ OHAUS 07 26 2017 05 16 37 Balance ID Balance Name PR2202 User Name Project Name Span Calibration Calibration is done Reference weight 2000 00 g Actual weight 2000 22 g Differen...

Page 33: ...ial glass cleaner Attention Do not use solvents harsh chemicals ammonia or abrasive cleaning agents 8 3 Troubleshooting TABLE 8 1 TROUBLESHOOTING Symptom Display Possible Cause Err 8 1 Over init zero...

Page 34: ...ing linearly to 50 relative humidity at 40 C x Operability is assured at ambient temperatures between 5 C and 40 C x Mains supply voltage fluctuations up to 10 of the nominal voltage x Installation ca...

Page 35: ...m 4 7 in 12 cm 4 7 in 12 cm 4 7 in 12 cm Span Calibration Points g 30 60 50 100 100 200 100 200 200 400 250 500 Linearity Calibration Points g 20 40 60 50 75 100 50 100 150 200 50 100 150 200 100 200...

Page 36: ...7 1 in 18 cm 7 1 in 18 cm 7 1 in 18 cm 7 1 in 18 cm 7 1 in 18 cm Span Calibration Points g 750 1500 1000 2000 2000 4000 2000 4000 3000 6000 Linearity Calibration Points g 500 1000 1500 500 1000 1500...

Page 37: ...nt weighing Platform Size diameter 3 5 in 9 cm 3 5 in 9 cm 4 7 in 12 cm 4 7 in 12 cm Span Calibration Points g 50 100 100 200 150 300 250 500 Linearity Calibration Points g 50 75 100 50 100 150 200 10...

Page 38: ...Percent weighing Platform Size diameter 7 1 in 18 cm Span Calibration Points g 400 800 1000 2000 2000 4000 2500 5000 1000 2000 2000 4000 3000 6000 Linearity Calibration Points g 200 400 600 800 500 1...

Page 39: ...PR Series Balance EN 37 9 2 Drawings and Dimensions Fully assembled dimensions Figure 9 1 0 001 g 0 0001 g model Figure 9 2 0 1 g 0 01 g model 11 93...

Page 40: ...stable non zero weight and stable zero reading to be printed xP Set Auto Print to Interval Print x print interval 1 3600 sec 0P disable the interval Print 0P 0P disable interval print continuous prin...

Page 41: ...ounting APW x in grams must have APW stored P Print Counting application APW x Set Percent application reference weight x in grams must have reference weight stored P Print Percent application referen...

Page 42: ...le Operating mode Full duplex Transmission mode Bit serial asynchronous Transmission code ASCII Baud rates 1200 2400 4800 9600 19200 38400 firmware selectable Bits parity 7 bit even 7 bit odd 7 bit no...

Page 43: ...he applicable weights and measures authorities If national regulations limit the validity period of the verification the user of the weighing instrument must strictly observe the re verification perio...

Page 44: ...been damaged by accident or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other than...

Page 45: ......

Page 46: ...ices worldwide Con oficinas en todo el mundo Avec des bureaux partout dans le monde Mit B ros weltweit Con uffici in tutto il mondo www ohaus com 30372570 P N 30372570 E 2018 Ohaus Corporation all rig...

Reviews: