OHAUS Navigator Calibration Manual Download Page 2

Navigator

 

Balance 

 

Calibration Guide

 

Navigator

 

 

Calibration Guide 

ii

 

Ohaus Corporation  www.ohaus.com 

LINEARITY CALIBRATION 

Linearity calibration uses three calibration points; zero, mid-scale and full scale.  

Before beginning calibration, be sure the menu Lock Switch is off.  

Note: 

Be careful not to touch the scale or the table while calibration is in progress, as it will 
cause the process to fail. 

 

5

 

  

(Release button) 

 

 

 

 

 

 

N

1

   

  

Y

1

 

 

 

 

 

Y

1

 

(On some models, no need  to press Yes)

 

(Place mass on platform)

 

 

 

 

 

Y

1

 

 

 

(Place mass on platform)

 

 

 

 

 

Y

1

 

 

 

 

 

 

 

 

 

 (Remove mass)

 

 

 

 

 

 

 

 

 

  

 
 
 

 

*83033982*

  

P/N 83033982

 