![OHAUS MC2000 Instruction Manual Download Page 10](http://html1.mh-extra.com/html/ohaus/mc2000/mc2000_instruction-manual_3617013010.webp)
EN-8 MC2000 Moisture Analyzer Series
1.
Please check points 3-8 in section 3.3.1 for detailed
information about the measurement process.
2.
The test order to build up a new category is from low
moisture level standard sample to high level (for example
test 15% level sample first, then 20% and then 25%...).
3.
We recommend the difference of moisture level between
each standard sample to be around 5%. And the whole
moisture range of these standard samples should be within
the real object range (for example 15%, 20% and 25%).
5. FACTORY RESET
1.
Enter category mode.
2.
Long press
button, till display shows
“-1-” and “C##”.
This function will delete all the adjustments and new categories
created by user.
6. MAINTENANCE
1.
After each test, clean the container with a brush.
2.
Please remove the batteries from analyzer in case of transit or storage.
7. GRAINMATE SOFTWARE
1.
By using GrainMate PC application software tool, the grain category
contents in MC2000 can be downloaded to PC and then transferred to
another unit by connecting the PC and MC2000 with USB cable.
2.
The GrainMate Software and instruction manual can be found here:
www.ohaus.com->SUPPORT->Software and Drivers.
8. SPECIFICATIONS
Model
MC2000
Moisture range (%)
3% - 45%
Readability (%)
0.1%
Repeatability (%)
0.2%
Sample volume
220 ml
Temperature range
0
℃
-50
℃
Display results
% Moisture, Weight, Density, Statistics
Power
4 AA Battery
Capacity of Category
50
Summary of Contents for MC2000
Page 2: ......
Page 24: ......
Page 36: ......
Page 48: ......
Page 60: ......
Page 63: ...MC 2000 Analyz tor vlhkosti pro zrna CZ 3 3 2 Displej...
Page 71: ...MC2000 RU 1 1 1 1 2 2 1 200 4 AA USB 2 2...
Page 72: ...RU 2 MC2000 2 3 a b 4 c 4 1 3 3 1...
Page 73: ...MC2000 RU 3 3 2...
Page 74: ...RU 4 MC2000 3 3 3 3 1 1 2 8 3 4 5 6 7 5 1 2 3 4 5...
Page 75: ...MC2000 RU 5 3 3 2 C 3 3 3 C 3 3 4 0 0 3 3 5...
Page 76: ...RU 6 MC2000 4 4 1 1 2 3 0 0 4 200 5 200 6 4 2 1 0 0 2 3...
Page 81: ...MC2000 RU 11 2002 96 EC WEEE 2008 2006 66 EC www ohaus com weee...
Page 82: ......
Page 83: ......