MB120 EN-37
Enter the time (5 seconds to 300 seconds).
Free (% / s)
Enter the weight loss (0.01% to 5.00%).
Enter the time (5 seconds to 300 seconds).
7.2.4 Display mode
Display mode (Display result) offers you the choice of whether your display indicates % MC, %
DC, % RG or grams during the measurement process.
Press the
Display mode
button.
Select %MC, %DC, %RG or g.
%𝑀𝐶(𝑀𝑜𝑖𝑠𝑡𝑢𝑟𝑒 𝐶𝑜𝑛𝑡𝑒𝑛𝑡) =
𝑆𝑡𝑎𝑟𝑡 𝑤𝑒𝑖𝑔ℎ𝑡 − 𝐷𝑟𝑦 𝑤𝑒𝑖𝑔ℎ𝑡
𝑆𝑡𝑎𝑟𝑡 𝑤𝑒𝑖𝑔ℎ𝑡
× 100%
%𝐷𝐶(𝑆𝑜𝑙𝑖𝑑𝑠 𝐶𝑜𝑛𝑡𝑒𝑛𝑡) =
𝐷𝑟𝑦 𝑤𝑒𝑖𝑔ℎ𝑡
𝑆𝑡𝑎𝑟𝑡 𝑤𝑒𝑖𝑔ℎ𝑡
× 100%
%𝑅𝐺(𝑅𝑒𝑔𝑎𝑖𝑛 𝐶𝑜𝑛𝑡𝑒𝑛𝑡) =
𝑆𝑡𝑎𝑟𝑡 𝑤𝑒𝑖𝑔ℎ𝑡 − 𝐷𝑟𝑦 𝑤𝑒𝑖𝑔ℎ𝑡
𝐷𝑟𝑦 𝑤𝑒𝑖𝑔ℎ𝑡
× 100%
Summary of Contents for MB120