9.1.4 Centrifuge FC5816
Model
FC5816
Speed Range
200 rpm -15000 rpm;10 rpm/set
Maximum RCF
21379 x g;10 x g/set
Maximum Capacity(Rotor)
6 x 250ml
Temperature range(N/A)
Air cool
Running Time
10 sec to 99 hr 59 min 59 sec or continuous
Noise level (depending on the rotor)
≤ 61 ± 2 dB(A)
Allowable density at maximum speed
1.2 g/ml
Allowable kinetic energy
34363 Nm
Mains power connection AC
230 V ~ 50/60 Hz
120 V ~ 50/60 Hz
Voltage fluctuation
± 10 %
± 10 %
Current consumption
2.4 A
4.2 A
Power consumption
530 W
520 W
Dimensions (W × D × H)
446 x 538 x 354 mm
17.6 x 21.2 x 13.9 in
Net Weight (without rotor)
52 kg
115 lb
Shipping Dimensions (W × D × H)
840 x 640 x 590 mm
33.1 x 25.2 x 23.2 in
Shipping Weight (without rotor)
77 kg
170 lb
Ambient conditions (EN/IEC 61010-1)
Environment
for indoor use only
Altitude
Use up to an altitude of 2000 m
Ambient temperature
2°C up to 35 °C
Max. relative humidity
Max. relative humidity 80 % for temperatures up to 31°C,
decreasing linearly to 50 % relative humidity up to 35°C.
Overvoltage category (IEC 60364-4-443)
II
Degree of contamination
2
Class of protection
I
Not suitable for use in hazardous environments.