38 MAINTENANCE
42
Short circuit in the frequency
converter
Frequency converter defective, motor
defective
43
Under voltage frequency converter
Brief power failure, At revision no.1 it could
be, that the frequency relay is defective,
cable break, plug is loosen (between
control board to frequency converter)
44
Over voltage frequency converter
Break resistant is defective, the rotor is
braked to strong
45
Over temperature frequency
converter
Motor runs hard, ventilation is defective,
frequency converter is not cooled enough
46
Over temperature motor
Motor runs hard, ventilation is defective
47
Over current frequency converter
Frequency converter is defective, motor is
defective
48
Timeout between control unit and
frequency converter
Software of the unit has a black out
49
Other error frequency converter
Summary of very specific error, which
should normally not occur.
(No release, internal timing error, system
error, reset by watch dog, start a trial with
direction of rotation error, program memory
CRC error)
55
Overspeed
Transmission error, firmware error,
frequency converter us defective.
70
Timeout at control unit through the
RS232 interface
RS232 connection is interrupted.
99
Rotor is not allowed in this
centrifuge
The number of the inserted rotor does not
exist in the data record of the software.
rotor no No rotor detected
It was no rotor installed. Transponder in
the rotor is defective or missing, antenna
defective, cable break, plug is loosen,
control board defective. In rare cases a
transponder-antenna combination can
cause a reduce range, that a conversation
can't occur.
FALSE The rotor, which is stored in the
program, does not match with the
inserted rotor.
There was chosen a program, in which a
certain rotor is stored, but the inserted rotor
is another one.