2.4 Installation the rotors 
FC5706 

FC5707, FC5513 

FC5515(R), FC5714, FC5716(R), FC5816(R), FC5916(R)

 

a.

Clean the drive shaft as well as the collet with a clean, grease-free piece of cloth.

 

 

 

b.

Place the rotor onto the drive shaft. Make sure that the rotor is fully installed onto the motor shaft

.

 

c.

Fasten the nut to hand-tight

FC5515 
Motor shaft and chamber 

a. Clean the drive shaft as

well as the collet with a

clean, grease-free piece of

cloth.

b.

Place the rotor onto the

drive shaft. Make sure

that the rotor is fully

installed onto the motor

shaft.

c.

Hold the rotor with one

hand and secure the rotor

to the shaft by turning the

fixing screw clockwise.

Rotor with separate nut

 

Tool for rotor with nut 

Tool for rotor without nut 

FC5515R 
Motor shaft and chamber 

FC5714/FC5718/FC5718R/
FC5816/FC5816R/FC5916/
FC5916R
Motor shaft and chamber 

Rotor with build-in nut

 

a. Clean the drive shaft as

well as the collet with a

clean, grease-free piece of

cloth.

b. Place the rotor onto the

drive shaft. Hold the rotor

with one hand and secure

the rotor to the shaft by

turning the fixing screw

clockwise.

Summary of Contents for Frontier FC5706

Page 1: ...Centrifuge User Guide For FC5706 FC5707 FC5513 FC5515 R FC5714 FC5718 R FC5816 R FC5916 R 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmen...

Page 2: ......

Page 3: ...to severe injuries or death if not avoided WARNING For a hazardous situation with medium risk possibly resulting in injuries or death if not avoided CAUTION For a hazardous situation with low risk re...

Page 4: ...uipment and inside the chamber Use precaution when moving or lifting the centrifuge as the equipment is heavy 2 1 Display Panel Control panel of different models are shown below FC5515R FC5718R FC5816...

Page 5: ...properly installed before each run Do not operate the centrifuge with rotors or buckets which show any signs of corrosion or mechanical damage Do not operate with extremely corrosive substances which...

Page 6: ...iece of cloth b Place the rotor onto the drive shaft Make sure that the rotor is fully installed onto the motor shaft c Hold the rotor with one hand and secure the rotor to the shaft by turning the fi...

Page 7: ...y be used if all positions are occupied with buckets and racks and the diagonally opposing pair must be buckets and racks of the same size and type Do not mix different types of buckets and racks diag...

Page 8: ...s before installing making connections or servicing this equipment Failure to comply with these warnings could result in personal injury and or property damage Retain instructions for future reference...

Page 9: ...personnel 3 6 Danger and precautions To protect people and environment the following precautions should be observed During centrifugation the presences of people are prohibited within 30 cm around the...

Page 10: ...ment WEEE this device may not be disposed of in domestic waste This also applies to countries outside the EU per their specific requirements Please dispose of this product in accordance with local reg...

Page 11: ...use harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio...

Reviews: