23
Mode
Units
Setup
Enter
4.2 Readout
The Readout menu is used to adapt the balance to environmental conditions. It contains four submenus: Stable,
Auto 0, Filter, Lock and Exit. Lock enables you to program balance parameters and to lock the settings.
Stability
The stability range specifies the weighing results and must be within a preset tolerance limit for a certain time to
turn the stability indicator ON. When a displayed weight changes beyond the allowable range, the stability indicator
turns OFF, indicating an unstable condition. Factory default setting is shown in bold type.
.5 d
Smallest range: stability indicator is ON only when displayed weight is stable within
.5 divisions.
1 d
Reduced range.
2 d
Normal range.
5 d
Largest range, stability indicator is ON even though displayed weight changes slightly.
When the RS232 interface is configured to print stable data only, the stability range also governs data output.
Displayed data will only be output if it is within the selected stability range.
Auto-Zero
Auto-Zero minimizes the effects of temperature changes and shift on the zero reading. The balance maintains the
zero display until the threshold is exceeded. Factory default setting is shown in bold type.
OFF
Turns Auto-Zero OFF.
.5 d
Sets threshold to .5 divisions.
1 d
Sets threshold to 1 division.
3 d
Sets threshold to 3 divisions.
Filter
Filter compensates for vibration or excessive air currents. Default settings are shown bold.
-0-
reduced stability, fastest stabilization time
-1-
normal stability, normal stabilization time
-2-
more stability, slow stabilization time.
-3-
maximum stability, slowest stabilization time.
Lock
Lock ON/OFF can only be changed when the hardware Lock Switch is set OFF/disabled. A menu is locked when
the menu lock is set ON and the Lock switch is ON. Lock when selected and turned on, locks all of the entries
made under the Readout menu. In the locked condition, items may be looked at but not changed in the menu.
When set off, entries may be changed. OFF is the default setting.
Setup
Use keys to change,
press Enter
Change
Procedure
To select any of the items in the Readout menu, pro-
ceed as follows:
• Press the Setup button, CAL is displayed.
• Press
or
button until READOUT is dis-
played.
• Press Enter button to save setting.
• Press
or
button until either STABLE, AUTO
0, FILTER, LOCK or EXIT is displayed.
• Press Enter button to save setting.
• Press
or
button and select the desired
menu setting.
• Press Enter button SAVED is displayed.
• Press
or
button to continue or EXIT.
• Press Enter button to save setting.