CHAPTER 4 TESTING
4.4 Consistency
Check
TABLE 4-2. LOAD CELL OUTPUT READINGS (in mV with 5V Excitation)
Model/Capacity
Zero Load
50% Load
100% Load
D15HR/15kg
1.3 ± 1.0
Zero + 2.8
Zero + 5.6
D30HR/30kg
1.1 ± 1.0
Zero + 2.55
Zero + 5.1
D60HR/60kg
0.7 ± 1.0
Zero + 3.1
Zero + 6.2
D60HL/60kg
0.7 ± 1.0
Zero + 3.1
Zero + 6.2
D100HR/100kg
1.3 ± 2.0
Zero + 2.7
Zero + 5.4
D150HR/150kg
0.53 ± 2.0
Zero + 2.5
Zero + 5
D300HL/300kg
0.34 ± 2.0
Zero + 3
Zero + 6
D10QR/10kg
4.3 ± 2.0
Zero + 2.3
Zero + 4.7
D25QR/25kg
1.6 ± 2.0
Zero + 2.8
Zero + 5.6
D50QR/50kg
1.7 ± 2.0
Zero + 2.4
Zero + 4.8
D100QL/100kg
1.3 ± 2.0
Zero + 2.7
Zero + 5.4
D250QX/250kg
0.34 ± 2.0
Zero + 2.5
Zero + 5
D10WR/10kg
4.3 ± 2.0
Zero + 2.3
Zero + 4.7
D25WR/25kg
1.6 ± 2.0
Zero + 2.8
Zero + 5.6
D50WR/50kg
1.7 ± 2.0
Zero + 2.4
Zero + 4.8
D100WL/100kg
1.3 ± 2.0
Zero + 2.7
Zero + 5.4
D250WX/250kg
0.34 ± 2.0
Zero + 2.5
Zero + 5
Defender™ H-Q-W Series Service Manual
4-3
Ohaus Corporation www.ohaus.com
Summary of Contents for Defender D100HL