
Defender 3000 Indicators
EN-13
2.3.2
EasyConnect
TM
Indicator
To connect an OHAUS EasyConnect
TM
scale base to an i-DT33P or an i-DT33XW EasyConnect
TM
indicator,
plug the base's connector into the external load cell connector located at the back (i-DT33P) or bottom (i-
DT33XW) of the indicator
. Then rotate the base connector’s locking ring in the direction of the arrows
demonstrated below.
Note
: For connecting bases which do not support EasyConnect
TM
to an EasyConnect
TM
indicator, contact an
authorized OHAUS dealer to obtain a load cell cable gland kit (P/N 30379716) as an accessory.
2.3.3
Non-EasyConnect
TM
Indicator
To connect a non-EasyConnect
TM
scale base to an i-DT33P or an i-DT33XW non-EasyConnect
TM
indicator,
follow the instruction in
Open the Housing
section to open the housing first before proceeding.
Note
: For connecting EasyConnect
TM
bases to a non-EasyConnect
TM
indicator, contact an authorized OHAUS
dealer to obtain a load cell cable gland kit (P/N 30427858) as an accessory.
2.3.3.1 i-DT33P
1.
After opening the housing, release the load cell connector (circled) on the rear of the indicator and then
pass the load cell cable through it.
Figure 2-1 i-DT33P
Figure 2-2 i-DT33XW
Summary of Contents for Dafender 3000 i-DT33P
Page 1: ...DefenderTM 3000 Indicators Instruction Manual i DT33P i DT33XW...
Page 2: ......
Page 59: ......