Switch On/Off

On

Off

Switch EC/TDS/SAL/RES Mode

EC 

TDS

 

SAL

 

RES

 

EC Reference Temperature Setup 

EC Temperature Coefficient Setup 

 Touch the value on the screen 

to confirm your setting. 

EC TDS Factor Setup 

 Touch the value on the screen 

to confirm your setting. 

EC Cell K Setup

 Touch the value on the screen 

to confirm your setting. 

Note

: Please set the Cell K value first before EC calibration. 

Calibration

Touch the 

Calibration 

icon on the 

screen.

Calibrating 

Press the 

 button when you 

finish to continue.

Touch 

Finish

 to complete 

calibration.

The calibration result will be 

shown when you finish. 

Note

: Predefined standard solution: 10 μS/cm, 84 μS/cm, 146.5 μS/cm, 500 μS/cm, 1413 μS/cm, 12.88 mS/cm. Please set the Standard Solution first before EC calibration. 

EC Standard Solution Setup

Temperature Unit Setup 

Temperature Setup 

 Touch the value on the screen 

to confirm your setting. 

Measurement

Insert the probe into a sample.

 

Measuring

Completed

Touch the mode area on the 

screen to select measurement 

mode. 

Store/Recall Data

Print:

Touch

 the 

Printer

 icon on 

the screen to print. 

Recall measurement data:

Touch

 the 

Beaker

 icon on the 

screen to recall. 

Note

: measurement data will 

be automatically stored after 

each operation. 

Print

 a-AB33EC Quick Start Guide