16
12.2.
Determination total hardness
The measurement can be aborted at any time using the
[◄]
key.
1. Open the menu selection using
[MENU].
2. Confirm “special functions” with
[OK].
3. Use
[
▲
]
or
[
▼
]
to change to “determination total hardness”
of the selected sensor and confirm with
[OK]
.
4. Put 20 ml of the sample in the sensor.
Dip the pH sensor in the water sample.
Start the measurement process with
[OK].
5. Measurement process: The sample is
measured (duration: 60 seconds).
6. Confirm measurement with
[OK].
7. Add 0.4 ml RM B. Insert the sensor in the sensor
vial, shake slowly several times. Start the measure-
ment process with
[OK].
8. Measurement process: The sample is measured
(Duration: 60 seconds).
9. Confirm measurement with
[OK].
10. Display total hardness.
Confirm measurement with
[OK].
pH value
21.3 °C
7.51
______________________________
The water sample is measured
Please wait!
60 sec
Confirm measurement with
[OK].
Cancel
[
◄]
pH value RM B
21.3 °C
6,27
______________________________
The water sample is measured
Please wait!
60 sec
Confirm measurement with
[OK].
Cancel
[
◄]
Main menu
special functions
determination
total hardness