In standalone mode, the test input data can be downloaded to the device to
perform off-line testing. After the input data is downloaded to the device, it
can be disconnected from the host PC and measurements can be performed
off-line. Once the measurement is completed, the device can be plugged
back into the PC and all test data will be downloaded from the device into the
PC and results analyzed automatically. The image below shows the confirma
-
tion dialog for settings updated on the device and also instructions on next
steps. Once the OK button is clicked, the software returns to the main screen.
In standalone mode, no analysis is performed on the device. All analyses are
performed on the PC. Multiple tests (100 max) can be performed in this mode
before results are downloaded to the host PC.
If multiple tests are performed, the same set of input data will be used for
all tests. If different input data are required for each test, the device must be
connected to the PC to update input data before each test.
1. Ensure the mud sample is homogenous. Mix the sample thoroughly using
a blender if necessary. Ensure there are no air bubbles in the sample be
-
fore inserting the probe. Immerse the probe so that the slots in the probe
shield are submerged.
2. Power on the device using the power button.
3.
Navigate to the “Measure” menu option using the navigation keys and use
the Enter key to select it.
4. The device will present the input parameters selected during set up.
Ensure these are correct and use the enter key to continue. Mud Density
can be adjusted on the device using the navigation and +/- keys.
Testing
Standalone