Specifications

Maximum Temperature: 500°F (260°C)

Maximum Pressure: 2,000 psi (13.8 MPa)

For tests above 200°F, refer to the chart below for the appropriate pres-

sure.

Do not use nitrous oxide cartridges as pressure sources for high-

temperature, high-pressure (HTHP) aging. Under high temperature and 

pressure, nitrous oxide can detonate in the presence of grease, oil, or 

carbonaceous materials. Nitrous oxide cartridges are to be used only 

for Garrett Gas Train Carbonate Analysis.

Carbon dioxide and nitrous oxide cartridges are pressurized to approxi-

mately 900 psi at 1 atmosphere (sea level). Therefore, they should never 

be transported by airplane without proper packing because cabin de-

pressurization may cause an explosion.

If the aging cells are going to be rolled in a roller oven during a test,  

install o-rings on the outer perimeter on the top and bottom of the cells. 

Failure to do so can damage the rollers in the oven. Teflon (#175-46), 

Viton

®

 (#175-47), and Buna N (#175-54) o-rings are available.

Cell Corrosion

Test fluids under high tempreature and pressure can corrode the cell body 

and caps. Carefully inspect the cell body and calls for corrosion before and 

after each test.

Some materials are more susceptible to corrosion than other. Also, some 

fluids and additives are more corrosive than others. OFITE offers a variety of 

cell materials for different levels of corrosion resistance and cost.

Mud Volume and Pressure for 

High-Temperature Aging

Aging 

Temp. 

(°F / °C)

Water 

Vapor 

Pressure 

(psi)

Coefficient 

of

Expansion 

of Water

Suggested 

Applied 

Pressure 

(psi / kPa)

Mud

Volume in

260 mL

Cell (mL)

Volume 

with 

Teflon 

Liner

Mud 

Volume 

in 500 

mL Cell 

(mL)

Volume 

with 

Teflon 

Liner

212 / 100

14.7

1.04

25 / 172

225

130

450

326

250 / 121

30

1.06

50 / 345

225

130

450

326

300 / 149

67

1.09

100 / 690

200

116

425

308

350 / 176

135

1.12

150 / 1,034

200

116

400

289

400 / 204

247

1.16

250 / 1,724

-

-

375

271

450 / 232

423

1.20

300 / 2,069

-

-

375

253

500 / 260

680

1.27

375 / 2,586

-

-

325

235

Summary of Contents for 175-65

Page 1: ...0 mL Double Capped 175 65 Instruction Manual Updated 6 18 2019 Ver 1 OFI Testing Equipment Inc 11302 Steeplecrest Dr Houston Texas 77065 U S A Tele 832 320 7300 Fax 713 880 9886 www ofite com Copyrigh...

Page 2: ...lecrest Dr Houston TX 77065 USA Tel 832 320 7300 Fax 713 880 9886 www ofite com 1 Intro 2 Components 3 Parts and Accessories 4 Specifications 5 Operation 6 Maintenance 8 Warranty and Return Policy 9 T...

Page 3: ...perature testing up to 500 F 260 C For prolonged exposure to elevated salinity at high temperatures such as 20 000 mg L chlorides at 350 F 176 6 C cells con structed from premium metals such as Hastel...

Page 4: ...Wrench for Set Screw 175 16 Valve Stem 175 47 O ring for Outside of Aging Cell Body Viton 90 Qty 2 175 65 001 Cell Body 316 Stainless Steel 175 65 002 Inner Cap without Pressure Port 316 Stainless Ste...

Page 5: ...o 400 F 170 17 Viton O ring for Valve Stem 175 09 2 Viton O ring for Inside of Aging Cell Body 175 47 Viton O ring for Outside of Aging Cell 175 62 Viton O ring for Teflon Liner Plug 175 63 Viton O ri...

Page 6: ...e outer perimeter on the top and bottom of the cells Failure to do so can damage the rollers in the oven Teflon 175 46 Viton 175 47 and Buna N 175 54 o rings are available Cell Corrosion Test fluids u...

Page 7: ...o ring so that it seats in place Hand tighten the outer cap in place Using the Allen Wrench tighten the set screws in the outer cap The ends of the cell body are identical The inner cap without the pr...

Page 8: ...m the valve stem or cell cap 7 Place the Aging Cell inside the oven and adjust to the desired tempera ture 8 After the desired aging time has elapsed remove the cell from the oven and allow it to air...

Page 9: ...se to the threads on the outer cap and cell body Every three years yearly for highly corrosive conditions 1 Sand blast the metal components cell body inner cap outer cap 2 Visually inspect the cell bo...

Page 10: ...ny losses or damages that occur as a result of Improper installation or maintenance of the products Misuse Neglect Adjustment by non authorized sources Improper environment Excessive or inadequate hea...

Reviews: