12
Quick Start
1.
Place the thermometer in the heating jacket.
2.
Preheat the heating jacket to 10°F (6°C) above the test temperature. The
pilot light will turn on at the target temperature.
3. Inspect all o-rings. Replace any that show signs of damage or wear.
4. Apply grease to all o-rings.
5.
Assemble both cell caps. Place an o-ring in the groove around the cap.
6. Make sure all threads are clean and free of debris.
7.
Apply anti seize compound (#165-44-2) to the threads of the cell locking
rings and valve stems.
8. Place an o-ring in the groove in the inlet side of the cell.
9.
Place cement screen on top of the o-ring.
10.
Screw the inlet cell cap into the cell and tighten it completely.
11.
Screw a valve stem into the inlet cell cap and tighten it completely.
12.
Pour the test fluid into the cell. Leave adequate space for fluid expansion.
13. Place an o-ring in the groove inside the cell.
14.
Place a cement screen on top of the o-ring.
15.
Screw the outlet cell cap into the cell. Tighten it completely.
16. Screw a valve stem into the outlet cell cap and tighten it completely.
17.
Invert the cell and place it in the heating jacket. Rotate the cell until it
seats.
18. Transfer the thermometer to the cell body.
19.
Connect the manifold block from the dual Nitrogen manifold to the top
valve stem and secure it with the safety pin.
20.
Screw the hose from the dual Nitrogen manifold into the back pressure
receiver.
21.
Connect the back pressure receiver to the bottom valve stem and secure
it with the safety pin.
22.
Connect a Nitrogen cylinder to the dual Nitrogen manifold.