HTHP Cell Assembly
The 170-12-2 HTHP cell assembly can handle maximum standard filtration
tests at elevated temperatures and pressures. The maximum pressure rating
is 2,000 psi and the temperature is 400°F (204°C). Most tests, conducted with
the 175 mL capacity filter presses under API standards, operate at an inlet
pressure of 600 psi and with temperatures up to 350°F (176°C). The cell bod
-
ies and cell caps are scribed with serial numbers so that the same cap and
same cell bodies are used on each and every test, resulting in more repro
-
ducible results. Additional information includes the part number, temperature
rating, the alloy and pressure rating of the cell.
All OFITE HTHP cell assemblies are made of Stainless Steel. For very corro
-
sive environments cell assemblies made of Hastelloy
TM
C-276 or Inconel are
available and recommended.