48
10. Fill the rotor cup with 200 cP calibration fluid and place it on the rotor.
11. Open the TransCal software supplied by OFITE.
12. On the “Manage Devices” screen, right-click the Viscometer in the list and
select “Default Device”. Then click “Done”.
If the Viscometer does not appear in the list, select the correct COM port
from the list and click the “Search for Devices” box.
13. Click the “Start Transducer Calibration” button. The transducer calibration
will run automatically for about 20 minutes. At the end of the procedure,
the status screen will disappear. Close the TransCal software and open
the ORCADA
®
software.
14. Calibrate the viscometer through the ORCADA
®
software before running
any other tests.