2021-03-23 / Rev. 10
VV016042
40
9.7.3.2
Calibrate PLD (Power Laser Diode)
Calibration of the PLD is performed on this page. Current power values are shown and new ones are entered here.
For GLM01, adapter
<Endo>
is used. Make sure the correct adapter is selected before calibration.
9.7.4
Short calibration guideline
– target laser TLD (aiming beam)
This chapter will guide you through a short calibration routine of the target laser diode.
Use a power meter, measurement unit WATT to perform measurement.
Perform measurement at the fibre connector at the front panel.
Procedure:
1. Start the laser in such a way that the aiming beam can be measured (
9.7.3.1, procedure 1
–6). The aiming beam is on
as soon as “Laser ON” is activated.
2. Measure the aiming beam at level 9 (
3. Change new value until a value of minimal 700µW is shown on the measurement device.
4. Enter a higher or a lower value; corresponding to the measured one. If the value is out of tolerance (700µW ± 300µW),
repeat 1-4.
(
5. Click on
<Write config.>
6. Change the value to 8 and back to 9 again (to ensure the setting change).