2021-03-23 / Rev. 10
VV016042
41
9.7.4.1
Calibrate TLD (Target Laser Diode)
Calibration of the TLD is performed on this page. The current power value is shown and the new one is entered here.
NOTE: The entered power value must be 40 or higher in order to achieve that the value of the aiming beam reaches the
specified value and tolerance range.
For GLM01, adapter
<Endo>
is used. Make sure the correct adapter is selected during calibration.
<Read Config>
reads the values from the module and displays them.
After calibration, the configuration must be written on the module by
<Write Config>
.
The configuration can be saved into a file by
<Export Config>
and restored by
<Import Config>
.
Summary of Contents for OS 4
Page 1: ......
Page 44: ...2021 03 23 Rev 10 VV016042 44 10 2 Core PCB ...
Page 134: ...2021 03 23 Rev 09 VV016042 134 18 Appendix C Safety check Please see form on next page ...
Page 136: ...2021 03 23 Rev 09 VV016042 136 19 Appendix D Functional test Please see form on next page ...
Page 151: ...VV016042 2021 03 23 Rev 10 ...