2021-03-23 / Rev. 10
VV016042
116
13.6
Sensors
13.6.1
Horizontal pedal sensor
VX310078 Pedal sensor horizontal
Quantity
Description
1
Pedal sensor horizontal
2
M3x12 oval head screw, TORX, ecosyn-fix
2
Cable tie
After the exchange of a pedal sensor, the pedal has to be sealed. For this purpose, additionally order
VX102479 sealing set
(see chapter 13.2)!
STEP 1:
1.
2.
3.
4.
Open pedal housing according to the
procedure described in
Remove battery pack according to
Disconnect cable from PCB.
Undo the grub screw.
STEP 2:
5.
6.
7.
Push out the wave vertical deflection
component from the base body.
Release spring.
Place rocker in such a way that ac-
cess is given to both sensors.
Do not lose the washers when
pushing out the wave vertical de-
flection and reassemble them
after repair.
Lubricate the wave vertical deflec-
tion before reassembling the pe-
dal.
Summary of Contents for OS 4
Page 1: ......
Page 44: ...2021 03 23 Rev 10 VV016042 44 10 2 Core PCB ...
Page 134: ...2021 03 23 Rev 09 VV016042 134 18 Appendix C Safety check Please see form on next page ...
Page 136: ...2021 03 23 Rev 09 VV016042 136 19 Appendix D Functional test Please see form on next page ...
Page 151: ...VV016042 2021 03 23 Rev 10 ...