Technical Data
GA 09.033/6.02 – (2011-07) –
©
Oerlikon Leybold Vacuum
11
Gauge supply
Voltage
+24 V=
±
5%
Current 750
mA
Fuse protection
900 mA with PTC element,
self-resetting after turning
CENTER ONE off or discon-
necting the gauge
Switching function
Number 1
Reaction delay
≤
10 ms if switching threshold
close to measurement value
(for larger differences con-
sider filter time constant)
Adjustment range
depending on gauge
(
→
[1] … [16])
Hysteresis
≥
1% FSr for linear gauges
≥
10% of measurement value
for logarithmic gauges
Switching function relay
Contact type
floating changeover contact
Load max.
125 V
∼
, 60 W (ohmic)
110 V=, 2 A, 60 W (ohmic)
DANGER
For benchtop use, max. 30 V~
or 60 V= may be connected.
Service life
mechanic
electric
10
8
10
5
(at maximum load)
Connector
9-pin D-Sub, male
(pin assignment
→
26)
Contact positions
2
3
4
Vacuum pressure
lower than switching
threshold
2
3
4
Vacuum pressure
higher than switch-
ing threshold or no
mains power