![Oerlikon Leybold Vacuum IONIVAC IE 414 Operating Instructions Manual Download Page 7](http://html1.mh-extra.com/html/oerlikon-leybold-vacuum/ionivac-ie-414/ionivac-ie-414_operating-instructions-manual_3397544007.webp)
Operating characteristics when used with control unit IM 540 and IM 520
IE 414
IE 514
Collector potential
0 V
Cathode potential
+ 80 V
+ 100 V
Anode potential
+ 220 V
+ 220 V
Reflector potential
--
205 V
Emission current range
0.06 - 0.6 mA
1.6 mA
Cathode heater current
1.5 A (typ.)
Cathode heater voltage
3 V (typ.)
3.7 V (typ.)
Sensitivity for nitrogen
17 mbar (typ.)
6.25 mbar (typ.)
Max. power when baking out
90 mA / 480 V
45 mA / 480 V
Description
7
300265097_002_A1 - 05/2014 - © Oerlikon Leybold Vacuum
69,4
33
DN 40 CF
66
25
12,8
98
58
12,8
DN 40 CF
55
90
69,4
33
Fig. 1 Dimensions in mm
IE 414