Trouble-shooting
OPM223/253
78
9
Trouble-shooting
9.1
Trouble-shooting instructions
The transmitter constantly monitors its functions itself. If an error occurs which the device
recognises, this is indicated on the display. The error number is under the unit display of the
main measured value. If more than one error occurs, you can call these up with the MINUS key.
Refer to the "System error messages" table for the possible error numbers and remedial
measures.
Should a malfunction occur without any transmitter error message, please refer to the
"Process-specific errors" or the "Device-specific errors" tables to localise and rectify the error.
These tables provide you with additional information on any spare parts required.
9.2
System error messages
You can display and select the error messages with the MINUS key.
Error
no.
Display
Tests and/or remedial measures
Alarm contact
Error current
Autom.
cleaning start
Facty User
Facty User
Facty User
E001
EEPROM memory error
1. Switch device off and then on again.
2. Load device software compatible with the
hardware (with optoscope, see
"Optoscope service tool" section).
3. Load measurement-parameter specific
device software.
4. If the error persists, send in the device for
repair to your local supplier or replace the
device.
Yes
No
—
—
1
E002
Instrument not calibrated,
calibration data invalid, no user
data invalid (EEPROM error),
instrument software not suitable
to hardware (controller)
Yes
No
—
—
1
E003
Download error
Invalid configuration. Repeat download, check
optoscope.
Yes
No
No
E004
Instrument software version not
compatible with module
hardware version
Load software compatible with hardware
Load measurement-parameter specific device
software.
Yes
No
No
E007
Transmitter malfunction,
instrument software not
compatible with transmitter
version
Yes
No
—
—
1
E008
SCS alarm:
Glass electrode: glass breakage
ISFET: leakage current > 400 nA
Check glass electrode for glass breakage or
hair-line cracks;
Inspect electrode plug-in head for moisture
and dry if necessary;
Check medium temperature.
Replace ISFET.
Yes
No
No
E010
Temperature sensor defective,
not connected or short-circuited
Check temperature sensor and connections;
check device and measuring cable with
temperature simulator if necessary.
Check correct option selected in field A5.
Yes
No
No
E030
SCS reference electrode warning
Check reference electrode for contamination
and damage; clean electrode.
No
No
No
E032
Slope range exceeded or below
range
Repeat calibration and renew buffer solution;
replace electrode if necessary, check device
and measuring cable with simulator.
Yes
No
—
—
1
E033
pH value zero too low or too high
Yes
No
—
—
1
E034
ORP offset range exceeded or
below range
Yes
No
—
—
1