Page 6 of 6
For Reference Only
TROUBLESHOOTING CHARTS
COMPACT 7600
IMAGE SYSTEM
3/12/97
IMAGE_C.DS4
VIDEO & VIDEO LEVEL
Approximate Voltages
(The Xray Head cover should be on for these measurements)
Chart 3
TEST PHANTOM
No object in the beam
Three 1mm copper filters
VIDEO SIGNAL
out of camera head
1.6V - 1.8 V Peak/Peak
1.6V - 1.8 V Peak/Peak
VIDEO SIGNAL
out of VD01 Pcb
0.8V - 1.0 V Peak/Peak
0.8V - 1.0 V Peak/Peak
DLIST SIGNAL
out of AU01 Pcb
6.0Vdc
6.0Vdc
kV
39 - 45
79 - 85
LED STATUS
CONDITION
Standby
Fluoro exposure
Loss of Camera Sync during Fluoro Exposure
No Camera Sync during Boot-Up
LED SV
ON
ON
ON
ON
LED L3
OFF
OFF
OFF
OFF
LED L2
ON
ON
ON
OFF
LED ERROR INDICATIONS DURING BOOT-UP
BMS 500 Frame Store Board Failures
ERROR CONDITIONS
CPU 100 Failure
CPU 100 EPROM/PROM Failure
CPU 100 RAM Failure
CPU 100 Clock Failure
BSA 500 Failure
BSA 500C Failure
BSA 500 RAM Failure
BSA 500C RAM Failure
BSA 500 IC Failure
BSA 500 Controller Failure
BSA 500 Failure
LED SV
OFF
LED 3
1 Blink
2 Blinks
3 Blinks
LED 2
1 Blink
2 Blinks
3 Blinks
4 Blinks
5 Blinks
6 Blinks
7 Blinks
IMAGE PROCESSOR FILTER CONTROL CODES
MODE
Standard Fluoro
0
0
H
0
H
0
0
H
H
0
0
0
H
H
H
H
H
H
0
H
0
0
0
H
H
0
0
H
FILTER CONTROL CODE
H
H
H
0
0
H
0
H
0
H
H
H
0
0
H
H
0
H
0
0
0
CONTROL LINES
Chart 4
DIAGNOSTICS
REMOVAL - REPLACEMENT