Ocutech VES

®

Mini

(Vision Enhancing System)

Fitting Guide

1. How to demonstrate the system: 

 

The Ocutech VES

®

 Mini is available in a demonstrator kit  

 

that uses a special adjustable frame. The kit can be ordered  

 

with one or two telescopes, though binocular demonstration  

 

is not recommended as the demonstrator frame mountings  

 

are not sturdy enough to maintain proper alignment. The  

 

two-telescope demonstrator kit is suggested as it allows for  

 

testing the system for either the right or left eye without  

 

having to disassemble the diagnostic telescope mounting.

 

 

 

 

 

When using the demonstrator frame, you can control the  

 

height of the telescope by raising and lowering the bridge  

 

piece. You can slide the telescope right and left to position  

 

the eyepiece directly above the pupil of the eye chosen to  

 

sight through the device. Confirm that the patient can see a  

 

full, round field through the telescope, and that the image  

 

is not cut-off to the right or left, or up or down. You may  

 

even want to allow the patient to slide the telescope slightly  

 

right or left themselves to position it best.

 

 

 

 

Once the position is finalized, remove the frame and  

 

measure the distance between the center of the eyepiece  

 

and the center of the bridge to determine the proper PD for ordering the telescope system. 

 

 

 

Focusing is accomplished by rotating the ring at the very front of the telescope. Patients occasionally try to  

 

turn the entire telescope to focus, rather than turning just the front ring. Warn them to NOT force the focus,  

 

as this may break the device. Do NOT use the PD markings on top of the demo frame. 

2. Indoctrinating the patient:

 

Once you have positioned the VES

®

 Mini properly, have the patient switch their fixation in and out of the  

 

telescope eyepiece until they get comfortable with the action. Demonstrate how to focus the device for  

 

different distances. 

 

 

 

Remind the patient that the VES

®

 Mini enlarges images three times and as a result they should be able to  

 

see targets three times further away than with their “normal” vision. Because of the “tunnel vision” field of  

 

view, they will need to learn to switch their attention from the regular lens (to find the target), to the   

 

telescope eyepiece (to see the target). They should practice this until they become comfortable and proficient. 

 

 

 

When looking at closer distances, they should be able to see print or items that are three times smaller than  

 

possible with their normal vision. Due to the shallow depth of field of telescopes at near, the patient can  

 

either move their head in and out to maintain clear vision at different distances or re-focus the device to  

 

maintain a clear image. 

 

Demonstrator frame
with one telescope installed. 

Focusing Ring 