6 Steam Sterilisation
16 / 26
Conditioning Instruction BIOM® 5 (G/55185/0818/en)
Pay particular attention to the following:
Smooth movement of the articulated joint
Function of the safety rod of the front loupe
Function of the focussing knob
Flip the BIOM® 5 towards the adaptor plate. Make
sure that the housing body can be shifted without
resistance.
Slide the safety rod of the loupe holder up and
down several times to check that it runs smoothly.
After the test, the rod must be pulled fully down to
the stop.
Check that the knob turns easily and shorten the
total length until the adjuster is at the uppermost
position.
Check whether all fastening screws are present
(e.g. screws at the feather key of the safety rod).
The control mark (1) must be below the guide.
Fig. 6–1: Test the safety functions
1
2
Control
mark
3