7 Use of the BIOM 4
Instruction manual BIOM 4 (G/54400/1011/en)
9 / 34
7
Use of the BIOM 4
7.1
Prior to each use
Under sterile conditions: Perform a safety function test
Make sure that all components are present and are sterile.
Flip the BIOM 4 towards the adapter plate. It is equipped with a swivel mechanism
for 90° and a detent for this purpose.
Make sure that the housing body can be swiveled without resistance.
fig. 7-1: Moving Components of the BIOM
Check that the safety slider for the lens holder runs smoothly by sliding it in and out
several times by hand.
Check that the focus adjustment wheel can be adjusted with ease, and shorten the
overall length of the BIOM until the bridge is located at the top position.
Attention
If one of these functions is not assured, the unit may not be used.
Before each use, check that
The unit is in technically perfect condition.
All connections and fasteners that can be loosened are properly tightened and are
in a safe condition.
The dovetail mount for the adapter is securely fastened at the micropscope.