![OCTELL OT-100 Operation Manual Download Page 12](http://html.mh-extra.com/html/octell/ot-100/ot-100_operation-manual_723819012.webp)
measured field strength of the product is higher than the applicable RF compliance
level described above, the product should be observed to verify normal operation.
If abnormal performance is observed, supplementary measures may be necessary, such
as reorienting or repositioning the product.
b In the entire frequency range from 150KHz to 80MHz, the field strength should be
lower than 3 V / m.
Recommended isolation distance between portable and mobile RF communication
equipment and this product
This product is intended for use in an electromagnetic environment where radio
frequency radiation disturbances are controlled. Depending on the maximum rated
output power of the communication equipment, the purchaser or user can prevent
electromagnetic interference by maintaining the minimum distance between portable
and mobile RF communication equipment (transmitters) and this product as recommended
below.
Transmitter rated
maximum output
Power / W
Transmitter's rated maximum output power / W
150 kHz ~ 80 MHz
d =
1.2
P
80 MHz ~ 800
MHz
d =
1.2
P
800 MHz~ 2.5 GHz
d =
2.3
P
0.01
Not applicable
0.12
0.23
0.1
Not applicable
0.38
0.73
1
Not applicable
1.2
2.3
10
Not applicable
3.8
7.3
100
Not applicable
12
23
For the transmitter's rated maximum output power not listed in the table above, the
recommended isolation distance d is in meters (m), and the corresponding transmission can be
used
The formula in the frequency column of the transmitter, where P is the maximum output power
of the transmitter provided by the transmitter manufacturer, in watts (W)
As a unit.
Note 1: At 80 MHz and 800 MHz, the higher frequency range formula is used.
NOTE 2 These guidelines may not be suitable for all situations. Electromagnetic propagation is
affected by absorption and reflection from buildings, objects and people.