Oxygen Analyzer/Monitor
OM-25-ME (or equivalent) Galvanic cell sensor, life
expectancy 2 years under normal conditions
Oxygen Analyzer Power Source
2 each AA batteries, life expectancy approx. 3000 use hours
Auxiliary O2 Flow Selector
Scaled 0-10 LPM in set increments, used for pre/post
anesthesia
Vaporizer
Penlon SigmaDelta Series, bolt (cage) mounted,
temperature compensated, very low maintenance
Airway Pressure Gauge
Dual scaled in cmH2O and mmHg, located on front
panel of ventilator
Pressure Gauge Tubing
May be attached to bag/vent switch arm or to a point
within the patient breathing circuit (operators choice).
Mechanical Ventilator
Pneumatically powered, time cycled, volume constant,
volume variable
• Ventilator Pressure Relief
Pre-set to maximum of 60 cmH2O located in main vent box
•
Pressure Relief Valve
Preset to 60 cmH2O
Ventilator Gas Power
40 to 70 psi, 50 psi optimal
Requirements
Use toggle switch to select gas source
Waste Gas Scavenger
Positive and negative relief valves, 1 L reservoir bag,
vacuum control knob
Total Machine Gas Leakage
@ 30 cmH2O
-0- ml/Min
@ 80 cmH2O
100 ml/min
I
nternal System Compliance
@ 20 cmH2O
1.1 ml/cmH2O
@ 40 cmH2O
1.3 ml/cmH2O
I
nternal System Resistance
@ 1.0 L/sec gas flow
4.11 cmH2O
@ 0.5 L/sec gas flow
1.80 cmH2O
APL Valve Pressure Drop
@ 3.0 L/min gas flow
0.12 cmH2O
@ 30 L/min gas flow
1.03 cmH2O
Oxygen Flush Valve
Recessed, self-closing, push-button, color coded and
labeled, provides 45-55 lpm constant flow, while
push-button is depressed; may be used sparingly with
O2 concentrator as O2 power source
CO2 Absorber System
King Systems KAB-9 (re-fillable) or KAB-1
(pre-filled/disposable)
CO2 Absorber Canister Capacity
400 grams soda lime
Directional Valves
Built in the CO2 Absorber
CO2 Absorber Holding Bracket
Plastic, secured with knob to main frame of machine
Bag-Ventilator Switch/PRV
Hand-operated selector switch and rotating knob for PRV
and Scavenger Outlet Port
and scavenging outlet
Bellows
Latex free, upward inflating, range from 0 to 1.6 L
Bellows Pressure Relief
Pre-set at 60 cmH2O
Common Gas Outlet
Quick-connect, size indexed
Tubing Circuit
King Systems F-360-61 or any standard anesthesia
circle circuit
Gas Pressure Hoses
DISS and thread indexed, female connectors at both ends
Gas Inlet Manifold
DISS and thread indexed, male connectors with one-way valves
Gas Inlet Manifold Filters
Located behind Manifold Air and O2 inlet male connectors
Gas Inlet Pressure Regulators
• Main supply cylinder
DISS/thread indexed for O2
• Safety back up cylinder
Pin-indexed, yoke mounted for "D" and "E" cylinders for O2
Oxygen and Air Supply Gauges
0-3000 psi range, color coded and clearly labeled
Oxygen Supply Alarms
• Main and safety back-up
Pneumatically actuated when O2 supply falls below 30psi
• O2 Concentrator
Pneumatically actuated when O2 supply falls below 1 psi
• Alarm power source
9-volt battery located in body of alarm box
• Alarm on/off
Labeled toggle switch located on body of alarm box
Air and O2 Flowmeters
Calibrated and scaled 0-10 lpm, color coded, O2
flowmeter has a fluted control knob for easy
identification by touch, alone
Oxygen Concentrator
To power O2 flowmeter only
Air Compressor
May be used to power ventilator and air flowmeter
14