Oxygen Analyzer/Monitor

OM-25-ME (or equivalent) Galvanic cell sensor, life
expectancy 2 years under normal conditions

Oxygen Analyzer Power Source

2 each AA batteries, life expectancy approx. 3000 use hours

Auxiliary O2 Flow Selector

Scaled 0-10 LPM in set increments, used for pre/post
anesthesia

Vaporizer

Penlon SigmaDelta Series, bolt (cage) mounted,
temperature compensated, very low maintenance

Airway Pressure Gauge

Dual scaled in cmH2O and mmHg, located on front
panel of ventilator

Pressure Gauge Tubing

May be attached to bag/vent switch arm or to a point
within the patient breathing circuit (operators choice).

Mechanical Ventilator

Pneumatically powered, time cycled, volume constant,
volume variable

• Ventilator Pressure Relief

Pre-set to maximum of 60 cmH2O located in main vent box

• 

Pressure Relief Valve

Preset to 60 cmH2O

Ventilator Gas Power

40 to 70 psi, 50 psi optimal

Requirements

Use toggle switch to select gas source

Waste Gas Scavenger

Positive and negative relief valves, 1 L reservoir bag,
vacuum control knob

Total Machine Gas Leakage

@ 30 cmH2O

-0-  ml/Min

@ 80 cmH2O

100 ml/min

I

nternal System Compliance

@ 20 cmH2O

1.1 ml/cmH2O

@ 40 cmH2O

1.3 ml/cmH2O

I

nternal System Resistance

@ 1.0 L/sec gas flow

4.11 cmH2O

@ 0.5 L/sec gas flow

1.80 cmH2O

APL Valve Pressure Drop

@ 3.0 L/min gas flow

0.12 cmH2O

@ 30 L/min gas flow

1.03 cmH2O

Oxygen Flush Valve

Recessed, self-closing, push-button, color coded and
labeled, provides 45-55 lpm constant flow, while
push-button is depressed; may be used sparingly with
O2 concentrator as O2 power source

CO2 Absorber System

King Systems KAB-9 (re-fillable) or KAB-1
(pre-filled/disposable)

CO2 Absorber Canister Capacity

400 grams soda lime

Directional Valves

Built in the CO2 Absorber

CO2 Absorber Holding Bracket

Plastic, secured with knob to main frame of machine

Bag-Ventilator Switch/PRV

Hand-operated selector switch and rotating knob for PRV

and Scavenger Outlet Port

and scavenging outlet

Bellows

Latex free, upward inflating, range from 0 to 1.6 L

Bellows Pressure Relief

Pre-set at 60 cmH2O

Common Gas Outlet

Quick-connect, size indexed

Tubing Circuit

King Systems F-360-61 or any standard anesthesia
circle circuit

Gas Pressure Hoses

DISS and thread indexed, female connectors at both ends

Gas Inlet Manifold

           DISS and thread indexed, male connectors with one-way valves

Gas Inlet Manifold Filters

Located behind Manifold Air and O2 inlet male connectors

Gas Inlet Pressure Regulators

• Main supply cylinder

DISS/thread indexed for O2

• Safety back up cylinder

Pin-indexed, yoke mounted for "D"  and "E" cylinders for O2

Oxygen and Air Supply Gauges

0-3000 psi range, color coded and clearly labeled

Oxygen Supply Alarms

• Main and safety back-up

Pneumatically actuated when O2 supply falls below 30psi

• O2 Concentrator

Pneumatically actuated when O2 supply falls below 1 psi

• Alarm power source

9-volt battery located in body of alarm box

• Alarm on/off

Labeled toggle switch located on body of alarm box

Air and O2 Flowmeters

Calibrated and scaled 0-10 lpm, color coded, O2
flowmeter has a fluted control knob for easy
identification by touch, alone

Oxygen Concentrator

To power O2 flowmeter only

Air Compressor

May be used to power ventilator and air flowmeter

14

Summary of Contents for Magellan 2200

Page 1: ...Magellan 2200 Model 2 Anesthesia Machine Service Manual 8005 Shannon Industrial Park Lane Atchison Kansas 66002 Tel 913 874 2000 Fax 913 874 2005 Email oceanicmedical lvnworth com Version 082206...

Page 2: ...xplodedView of Ventilator Control Knobs 11 Testing After Storage 12 Re Supply Component Kits and Installation Intervals 12 Maintenance Components Kits and Installation Intervals 12 APPENDIX 13 Specifi...

Page 3: ...el 2 Anesthesia Machine is tested utilizing certain testing equipment which may or may not be available in the field or in a general biomedical maintenance facility This manual will explain how the Op...

Page 4: ...r pliers D 1 sealing tape roll E Tube of Krytox or other oxygen compliant lubricant F Water based leak testing solution ANNUAL MAINTENANCE PARTS NEEDED ANDTESTING Recommended Inspection Replacement of...

Page 5: ...o the Gas Manifold with original or new filter E Install the filter spacer aluminum ring F Pull open the one way valve located on the end of the threaded portion of the DISS adapter and apply some Kry...

Page 6: ...y forward from the main frame B Removal of Rubber Bellows from Bellows Base Inspection of ExhalationValve DiaphragmValve and Installing New Bellows 1 Turn the clear plastic bellows canister cover coun...

Page 7: ...remove any contamination from the surfaces of both components NOTE Any anesthesia breathing circuit filter may be used After cleaning check the small O Ring C located in the bellows base under the dia...

Page 8: ...rmly into the receptacle then pulling the tubing backwards firmly C Push the ventilator bellows system back into its normal position D Re install the two ventilator base knob screws from the backside...

Page 9: ...r Position on theVentilator Gas SelectorToggle Switch C Air Flowmeter II ROUTINE FUNCTIONALTEST The functional test may be done at annual intervals or at any time that internal componenets have been r...

Page 10: ...s of air or oxygen Appropriate hose set to connect to gas source Pre Use Checklist OperatingVentilator Test Lung PatientTubing Circuit Procedure Attach the patient circuit to the equipment and to the...

Page 11: ...st lung utilized Reservoir bag and rubber test lungs may not inflate fully or over inflate causing erroneous results III TESTING AFTER STORAGE The following procedures should be used after the Magella...

Page 12: ...interval without installing the complete Kit A 108 1 Bellows Latex free A 108 2 Tubing Kit with valves A 108 3 Vaporizer End Caps A 108 4 Manifold Filters and Spacers A 108 5 Tweezers for removing Fil...

Page 13: ...20 4 kg 45 lbs Machine Materials Aluminum brass and plastic CASE DIMENSIONS Height 66 0 cm 23 inches Depth 60 96 cm 24 inches Width 53 34 cm 17 inches Weight 121 2 kg 55 lbs empty 200 5 kg 100 lbs ma...

Page 14: ...ly with O2 concentrator as O2 power source CO2 Absorber System King Systems KAB 9 re fillable or KAB 1 pre filled disposable CO2 Absorber Canister Capacity 400 grams soda lime DirectionalValves Built...

Page 15: ...er Check that opposing Air or O2 flowmeter is OFF Ventilator does not cycle Check Air or O2 gas power supply for 40 70 PSI on pressure gauges Check Vent Power Selector Switch in Air or O2 position Che...

Page 16: ...each 4 11 3 Bronze Filter cone shaped 2 each 4 2 2 Hose Connector Locking Straight 2 each 4 PART LIST FOR MAGELLAN 2200 MODEL 1 Listed by Assembly Section in BoldType NOTE Some numbers are repeated as...

Page 17: ...d Screw 8 Ops 16 0 Scavenger Waste Gas 11 Ops 16 1 ScavengerTubing 3 ea and TEE Adapter 11 Ops 16 2 Reservoir Bag 1 Liter 11 Ops PART NO NSN DESCRIPTION PAGE NO 17 0 Bulb Squeeze Hi Lo Pressure Check...

Page 18: ...nnection 7 Ops A 103 Carrying Case Air Compressor A 104 Carrying Case Magellan 2200 Anesthesia Machine 5 Ops A 105 Oxygen Cylinder Mount for Carrying Case 5 Ops A 106 Aluminum Trolley for Magellan 220...

Page 19: ...MAGELLAN 2200 ANESTHESIA MACHINE FRONT 19...

Page 20: ...MAGELLAN 2200 ANESTHESIA MACHINE BACK 20...

Page 21: ...GAS FLOW DIAGRAMS 21...

Page 22: ...22...

Page 23: ...23...

Page 24: ...24...

Page 25: ...25...

Reviews: