Ocean Optics LS-1-CAL Operating Instructions Download Page 2

6.

Find the on/off switch on the back of the lamp and turn the lamp on.

7.

Let the lamp warm up for at least 15 minutes before using.

8.

Insert the disk that came with your lamp into your computer. The disk contains two ASCII files. These files have
the same information as the Lamp Calibration Reports that came with your LS-1-CAL. One file has the
calibration numbers for calibrating the spectral response of your system with the lamp and a 

bare fiber

; its name

contains the lamp’s serial number, followed by 

FIB.LMP

. The second file has the calibration numbers when

calibrating the spectral response of your system with the lamp and a 

CC-3 cosine-corrector and fiber

; its name

contains the lamp’s serial number, followed by 

CC3.LMP

.

9.

Copy these two files into your OOIIrrad Irradiance Software directory.

10. Start OOIIrrad Irradiance Software. Make sure you are in 

Scope Mode

.

11. Under the 

Lamp

menu options, choose 

Select Lamp

. A window opens in which you must choose the Lamp

Calibration Report file that reflects your optical setup.

12. Under the 

Spectrometer

menu options, choose 

Configure Fibers

. Enter the fiber diameter values for each

channel in your setup. If using a bare fiber, enter the fiber’s diameter here. If using a fiber with a CC-3 cosine
corrector, enter 3,900. Whatever optical setup you wish to use for your application, you must also use for
calibrating the spectral response of your system. For example, if you are going to use a 200-

µ

m fiber with a CC-3

cosine corrector for your application, you must use the same 200-

µ

m fiber and CC-3 for calibrating the spectral

response of your system.

13. To the right of the displayed spectrum, enter the data acquisition parameters for your setup. The values for Scans

to Average and Smoothing Size must be the same for your reference, dark and irradiance scans. You may use a
different integration period for your reference scan, but you must use the same integration period for your dark
and irradiance scans.

14. Under 

Spectrometer

menu options, choose 

Calibrate

and then select the channel you are calibrating.

15. A dialog box opens with the message 

Verify lamp was ON for at least 15 minutes for a REFERENCE scan.

Make sure nothing is blocking the light path going to your spectrometer. Your setup (spectrometer, fiber, and
sampling optics) for taking a reference should be identical to your setup for measuring your sample. Taking a
reference spectrum is requisite before the software can calculate absolute irradiance measurements. Click 

OK

.

16. Another dialog box opens with the message 

Block light path to spectrometer for a DARK scan.

If possible, do

not turn off the light source. Instead, completely block the light path going to your spectrometer by inserting an
opaque object into the lamp’s filter slot. Taking a dark spectrum is requisite before the software can calculate
absolute irradiance measurements. Click 

OK

. The spectral response of your system is calibrated.

17. If you selected 

Display calibration info when calculating

in the 

Configure Spectrometer

dialog box, you will

now see the results of the calibration procedure. The top left graph represents the linear 15

th

order polynomial

regression. The top right graph represents the dark scan. The bottom graph is the calibration curve representing
the spectral response of the process.

18. The calibration procedure is saved in a file named 

CH0.CAL

for calibrating the master spectrometer channel,

CH1.CAL

for calibrating the first slave spectrometer channel, and so on.

Specifications

Operating Instructions: LS-1-CAL

- 2 -

Spectral range (calibrated):

300-1050 nm

Dimensions:

9.0 cm x 5.0 cm x 3.2 cm (LWH); 3.5" x 2.0" x 1.25" (LWH)

Power input:

12 VDC/800 mA (regulated)

Power output:

6.5 watts

Bulb life:

900 hours (recalibrate after ~50 hours of use)

Bulb color temperature:

3100K

Output to bulb:

5 volts/1.3 amps

Output regulation:

0.2% voltage

Time to stabilized output:

~15 minutes

Connector:

SMA 905