Ocean Optics HL-2000-HP-232 Installation And Operation Manual Download Page 18

A: ASCII Commands

10

000-10000-140-02-1209

Command

Function

Description

Example

DI

Disable Drive

Execution of this command disables the drive 
electronics and removes all current from the 
motor

DI

EN

Enable Drive

This instruction enables the drive electronics 
and allows the servo controller to become 
active, providing motor current as required.

EN

LL

Load Position Range 
Limits

The range limit mechanism provides an 
additional layer of protection. Both positive and 
negative range limits are determined 
independently using the sign of the limit 
parameter. If the controller receives an 
instruction from the host to move to a position 
that is not within the range limit window, it will 
stop at the set limit position. 

LL 2000000 
or

LL -5000000

Configuration-Related Commands

Command

Function

Description

Example

POR

Set Proportional Loop 
Gain

The proportional gain determines the system's 
proportional response to a given amount of 
positional error. Increasing this parameter 
provides a tighter and more dynamically 
responsive system. (Allowed value 0 – 255)

POR 20

I

Set Integral Loop Gain

Unlike the proportional gain, where the 
response remains constant if there is no 
change in error, the integral term continues to 
increase the effects of its response until it 
becomes effective. This parameter determines 
the rate of change of this response.

(Allowed value 0 – 255)

I 10

LPC

Load Peak Current 
Limit

This command sets the Peak current limit. The 
LPC argument is in milliamperes.

LPC 1000

LCC

Load Continuous 
Current Limit

This command sets the continuous current limit. 
The LCC argument is in milliamperes. 

LCC 500

RN

Reset Node

This command resets the gains and current 
limits to the values stored in the controller ROM 
(the defaults).

RN

Summary of Contents for HL-2000-HP-232

Page 1: ...as Ave Dunedin FL USA 34698 Phone 727 733 2447 Fax 727 733 3962 8 a m 8 p m Mon Thu 8 a m 6 p m Fri EST Ocean Optics Mikropack Maybachstra e 11 D73760 Ostfildern Germany Phone 49 0 711 34 16 96 0 Fax...

Page 2: ...onsent of Ocean Optics Inc in any form of binding or cover other than that in which it is published Trademarks Microsoft Windows Windows 95 Windows 98 Windows Me Windows NT Windows 2000 Windows XP and...

Page 3: ...t This instrument should not be used for any Clinical or Diagnostic purposes Data generated in these areas is not warranted in any way by Mikropack GmbH Any defects covered by this Warranty shall be c...

Page 4: ...Important Safety Notices B 000 10000 140 02 1209...

Page 5: ...Overview 1 Setting Up the HL 2000 HP 232 Light Source 1 Unpacking the Light Source 1 Package Contents 2 Set Up 2 Chapter 2 Operation 5 Optimizing the Optical Power Output 5 Shutter Diagram 6 Chapter 3...

Page 6: ...Table of Contents ii 000 10000 140 02 1209 Appendix B HL 2000 HP 232 Specifications 15 Specifications 15 Parts List 16 Index 17...

Page 7: ...instructions for replacing the bulb Appendix A ASCII Commands Contains a list of the ASCII command set used to control the light source s attenuator and shutter Appendix B HL 2000 HP 232 Specificatio...

Page 8: ...that you need Ocean Optics to make a change or an upgrade to your system To facilitate these changes you must first contact Customer Support and obtain a Return Merchandise Authorization RMA number P...

Page 9: ...requiring large diameter optical fibers or fiber and probe bundles The attenuator and shutter can be controlled by a series of ASCII commands issued by a host computer via RS 232 Setting Up the HL 20...

Page 10: ...3 Use this instrument in a clean laboratory environment Package Contents Your package should contain the following One HL 2000 HP 232 Light Source One serial cable One 24 VDC 1 2A power supply HL 200...

Page 11: ...lamp using the power switch on the front of the light source The shutter opens and the lamp flashes once Notes The light can only be turned on via RS232 see To test the light source use the HL 2000 HP...

Page 12: ...1 Setup 4 000 10000 140 02 1209...

Page 13: ...ed or a different fiber bundle diameter is used you can adjust the optical power of the unit Procedure Follow the steps below to adjust the optical power of the HL 2000 HP 232 Light Source 1 Connect a...

Page 14: ...2 Operation 6 000 10000 140 02 1209 3 Shift the SMA socket to optimize the optical power of the light source 4 Tighten the blocking screw to secure the SMA socket position Shutter Diagram...

Page 15: ...he bulb in the light source disconnect the lamp from your power source and allow the unit to cool for at least twenty minutes if necessary Bulb Replacement Procedure Procedure 1 Remove the 19 inch cas...

Page 16: ...10000 140 02 1209 4 Loosen the hex screw on the lamp housing 5 Remove the bulb by pulling it back away from the connector 6 Replace the spare bulb with a new one 7 Replace the hex screw on the lamp ho...

Page 17: ...n relative to the present position LR 10000 SP Load Command Velocity This parameter should be loaded with the desired maximum profile velocity in position mode This value should be selected to not exc...

Page 18: ...Commands Command Function Description Example POR Set Proportional Loop Gain The proportional gain determines the system s proportional response to a given amount of positional error Increasing this p...

Page 19: ...mum PWM signal SOR2 and the Maximum Velocity for 10 V at the analog input SOR1 SOR 1 EEPSAV Save To EEPROM This command saves the current parameters and configurations to EEPROM The saved parameters a...

Page 20: ...due to HL HN and HA commands HP 0 HS External Switch Status This command returns the state of the limit switch If the limit switch was active and an HS request is sent the controller will return a 1 I...

Page 21: ...via RS232 Bit 2 1 Speed command is PWM SOR 2 0 Speed command is analog voltage SOR1 Bit 3 1 Amplifier Enabled 0 Amplifier Disabled Bit 4 1 In Position 0 Not in Position Bit 5 1 Rising edge on externa...

Page 22: ...Commands 14 000 10000 140 02 1209 Command Function Description Example VER Get Version This command returns firmware version VER GPOR Get Proportional Term This command returns Proportional gain setti...

Page 23: ...ts list Specifications Specification Criteria Dimensions 70 mm x 100 mm x 160 mm Weight 600 g Pin description Pin 2 RX Pin 3 TX Pin 5 Ground Output 20 watts Output to bulb 1 6 A 24 VDC Wavelength rang...

Page 24: ...232 Specifications 16 000 10000 140 02 1209 Parts List Spare Parts Order Information Item Number Halogen high power 20 W light source with RS232 control HL 2000 HP 232 Halogen spare bulb High Power H...

Page 25: ...al switch limit configuration 11 motion related 9 D document audience iii purpose iii summary iii O optical power optimizing output 5 P package contents 2 parts list 16 product related documentation i...

Page 26: ...Index 18 000 10000 140 02 1209...

Reviews: