A: Maintenance
14
000-10000-025-02-201510
WARNING
Before replacing the bulb in the DH-2000, disconnect the lamp from your power
source and allow the unit to cool for at least twenty minutes, if necessary.
Do not touch the lamp glass directly, as contact with bare fingers will reduce the
lifetime of the bulb.
Replacing the Deuterium Bulb for All Models Except DH-
2000-BAL
►
Procedure
1.
Open the six slotted screws (14) and open the casing cover.
2.
Open the screws (15) with the tool (18) that is delivered with the spare bulb (16).
3.
Disconnect the old bulb and connect the new Deuterium Tungsten-Halogen lamp only with the
originally supplied connection plugs (17).
Replacing the Deuterium Bulb for DH-2000-BAL Model
►
Procedure
1.
Open the six slotted screws (14) and open the casing cover and rotate or remove the cover to
expose the bulb housing.