short transients and so give poorer detection of delphinids) or an intermediate value which will tend 

to reduce sensitivity to both short and long transients but not by much.

 

Blanking time:

 Time that must then elapse after a detection before another detection can be made. 

Useful to avoid detecting echoes from a click or to prevent detecting a large number of clicks if an 

animal is buzzing near the recorder.

 

Store Snippets: 

Select to enable storage of detection snippets.  

Snippet pre- and post-trigger durations

: The time with respect to the detection at which the snippet 

will be taken. The snippet starts 

pre-trigger

 seconds before the detection and ends 

post-trigger

 

seconds after the detection. These are both limited to 1.5 ms maximum which is more than long 

enough to cover odontocete clicks. The number of samples saved per snippet is determined by the 

sum of the pre- and post-trigger times. With the maximum settings, 1728 samples will be saved per 

snippet. 

 

4.1.4

 

Output File formats 

The ST click detector produces two additional output files per recording with suffixes 'bcl' and 'dwv'. 

These are in addition to the .wav and .xml files generated normally. The Pamguard interface 

automatically interprets these files when displaying detection and continuous recording results. For 

users interested in accessing the data using other interfaces, the details of the file formats are given 

below: 

 

xml

 

This is the metadata file produced by SoundTrap to document each recording. The 

ST detector adds metadata about the detector and the user settings to this file.  

 

wav

 

This is a Microsoft WAV format file containing the decimated continuous sound 

recording produced in parallel with the detector. The sampling rate is selected by the user 

but would typically be 48kHz or 96kHz to maximize recording time. By default, sound data is 

compressed in the SoundTrap using the lossless X3 compressor but this is transparent to the 

user: the .wav file contains the reconstituted uncompressed data. 

 

dwv

 

This is a Microsoft WAV format file containing the wideband sound snippets for each 

detection. The sampling rate is 576 kHz and the snippets are arranged contiguously with a 

constant number of samples per snippet. The size of each snippet is set the by the pre- and 

post-trigger user parameters. The snippet size in samples will be close to 

round((pre+post)*576e3) but the precise sample count can be found in the .xml file (see the 

Matlab tools for an example of how to extract this information automatically). By default, 

Summary of Contents for SoundTrap ST300 Series

Page 1: ...SoundTrap ST300 and ST4300 series user guide September 2021...

Page 2: ...g the IR Remote Control 9 2 6 Data Offload 10 3 More details 12 3 1 Data Files 12 3 2 Battery charging 12 3 3 Using the optional external battery pack 14 3 4 Calculation of deployment endurance 15 3 5...

Page 3: ...shooting 24 4 Appendices 26 4 1 SoundTrap HF Click Detector 26 4 1 1 Click detector 26 4 1 2 Snippet extractor 27 4 1 3 User parameters 27 4 1 4 Output File formats 28 4 1 5 Click Detector Caveats and...

Page 4: ...tallation and basic deployment instructions This is followed by several sections containing more detailed information around operation deployment maintenance and troubleshooting It is highly recommend...

Page 5: ...lation o The installation will be smoother if you re logged into windows as an administrator Visit www OceanInstruments co nz downloads to download the software Download and run the installer Once the...

Page 6: ...Figure 1 The Device List 2 3 Check SoundTrap Status Note the Status panel on the right Figure 3 Verify that the battery has sufficient charge and the memory is not full...

Page 7: ...pre determined date and time or manually using the IR remote control Note that the remote can always be used to override the other options o The Recording Schedule section provides selection of conti...

Page 8: ...boat Turning this on reduces sensitivity in the band 0 to 600 Hz while leaving higher frequencies unaffected o Detector see HF Click Detector section below o Ancillary sensors select those you want t...

Page 9: ...a later time the recorder will blink red to indicate it is armed and waiting The armed state is cancelled by pressing Stop key Re enable the armed state by pressing the ARM button You can verify reco...

Page 10: ...and whether it has yet been downloaded to your computer s hard drive Download individual files by right clicking the file and selecting download By default all files will be downloaded to a SoundTrap...

Page 11: ...e un check the Decompress checkbox This will disable the decompression step of the offload process meaning you will end up with sud files only These can be decompressed into wav files at a later time...

Page 12: ...equates to 1 hour recordings for a 288khz sample rate Where smaller than 2G files are desired for continuous recording it is recommended to set a duty cycle with zero off time For example a duty cycl...

Page 13: ...o not start the Host software or alternatively charge the SoundTrap from a USB power supply rather than a computer The battery should not be left for long periods in a discharged state When storing a...

Page 14: ...e o ring and the tubes sealing surface Close the housing by screwing closed by hand until the shoulder snugs up against the titanium tube You may choose to use a lever arm e g screwdriver through the...

Page 15: ...ple a SoundTrap 202 has 64 GB of memory dividing this by the minimum file size of 4 MB gives a maximum of 16000 files This limitation becomes an issue when using low sample rates in combination with X...

Page 16: ...of SD flash memory cards is such that occasional the memory card will fail to keep up with sampling thereby resulting in dropped samples When this occurs an entry will be written to the log xml files...

Page 17: ...ild up particularly when connected to USB and recharging the battery Accelerometer data is recorded as unscaled 16 bit values ranging 32768 The full scale sensitivity of the accelerometer is 2 g meani...

Page 18: ...onsumption and data rate We recommend using our online endurance calculator to make these calculations Note that the calculator assumes use of HTI 96 MIN hydrophones which consume approximately 3 6 mW...

Page 19: ...rcumference provide attachment points for cable ties for this purpose Do not use hose clamps or any other fastener that will place clamping stress on the housing as this may cause damage There are als...

Page 20: ...ndow which over time will tend to grow in size Standard housing 500m rated Most ST300 recorders are of this type For this housing topping up is only necessary when deploying deeper than 200m If the ai...

Page 21: ...e reset battery you should connect the 9V battery to the battery connector provided on the USB cable Note that you must make this connection in one smooth motion being careful not to bounce the connec...

Page 22: ...3 3 value from calibration sheet cal power 10 cal 20 convert calibration from dB into ratio y y cal multiply wav data by calibration to convert to units of uPa PAMGUARD PAMGAURD expects calibration da...

Page 23: ...ecommend keeping them enabled where possible as they provide a useful check of system calibration and performance ST4300 Calibration ST4300 recorders have a fixed gain of 4dB Overall system calibratio...

Page 24: ...he remote 5 The blue LED indicator should now light to indicate battery is charging 6 Leave for at least 30 minutes to allow the battery to recover 7 Start the host software and confirm device connect...

Page 25: ...charge Solution This most often occurs in cases of complete battery discharge followed by a long period of disuse For future avoidance of this problem ensure that you store SoundTraps with the battery...

Page 26: ...configured to also make a continuous recording at lower bandwidth that allows you to quantify the general ambient noise conditions an important factor affecting detector performance as well as the pre...

Page 27: ...important to note that the snippet is taken from the whitened sound and so is high pass filtered 4 1 3 User parameters The user parameters can be set by clicking the Configure button in the Detector...

Page 28: ...ormally The Pamguard interface automatically interprets these files when displaying detection and continuous recording results For users interested in accessing the data using other interfaces the det...

Page 29: ...0 and the noise level at the time of the detection Fields Info3 and Info4 are not used in detection reports The second report type is an effort report which has a type of E The effort report indicate...

Page 30: ...warranty implied or otherwise nor implication of fitness for any particular purpose Please use responsibly Test that the code functions correctly on your SoundTrap before deploying it e g by setting t...

Reviews: