4.6 Electrical installation
4.6.1 Connections
The OMC-160 anemometer is provided with a junction box to connect the sensor to the field cabling.
The junction box provides a M20 gland for the power & signal cable.
If the OMC-160-3 heating option is fitted, an additional M20 gland is provided for a separate cable for
24 V heater power supply.
Table 1
:
Connector X1
Recommended cable: shielded 2x2x0.75mm2 (like OMC-494)
Terminal 5
must be used for NMEA shield connection.
Terminal 8
is meant as 0 reference in case the receiver has a RS422 input, a separate wire should be
used and connected to the signal ground of the receiver.
Do not use terminal 8 for shield connection, this can cause an earth fault!
Terminal
Function
Remarks
1
Power GND
2
Power +
3
NMEA output A / RS422 -
4
NMEA output B / RS422 +
5
Ground (connected to housing)
Shield connection NMEA out
6
Heater power, 24V (AC or DC)
Optional
7
Heater power, 24V (AC or DC)
Optional
8
RS422 0
Use only for RS422