DIReCTIOnS fOR USe
PROCeDURe ChAIR fACe ReST CReSCenT
The Procedure Chair Face Rest Crescent is designed to be used for prone positioning. An Adjustable Head
Rest is required for this accessory.
T-RAIl ADAPTeR
The T-Rail Adapter can be attached at designated locations on one or both sides of the back rest sec-
tion, allowing the use of any t-rail accessory. Attach or remove using knobs. The T-Rail Adapter has a
Maximum Load Capacity of 150 lbs. (68 kg.)
POWeR OUTleT USe
The maximum amperage for the Power Outlet is 3 amps. Connecting equipment that exceeds this
rating will damage the system. Only connect low current devices. Do not connect an extension
cord or power strip to the Power Outlet. Do not touch equipment connected to the Power Outlet and the patient at the same
time. Any electrical accessories used with this table must comply with approved medical standards. Equipment relying only on
Basic Insulation (only one means of electrical protection for the patient / pratitioner) must not be used.
WARnInG
The Power Outlet can be used to connect low
power devices such as localized illumination,
computers, micro-controllers, etc. Exceeding the
maximum current will interrupt the power.