- 15 -
17. FIELD CALIBRATION LOCKOUT
AND RE-ENABLE
The calibration lockout feature prevents
any field calibration changes. The lockout
remains in effect until a lockout re-
enable has been performed. Use the
following procedures to lockout or re-
enable the field calibration operation.
LOCKOUT PROCEDURE
1. Turn the thermometer off.
2. Simultaneously press and hold the
cal
and the
°C/°F
keys down and
momentarily press the
on/off
key.
Continue to hold the
cal
and
°C/°F
keys
until the measurement mode is
displayed.
RE-ENABLE PROCEDURE
1. Turn the thermometer off.
2. Simultaneously press and hold the
hold
and the
cal
keys down and
momentarily press the
on/off
key.
Continue to press the
hold
and
cal
keys
until the measurement mode is displayed.